M I L L E N N I U M M E D I C A L P R O D U C T S , I N C .
18
SYMPTOMS
POSSIBLE CAUSE
CORRECTIVE ACTION
1) No motor sound or movement of
piston rod
•
Actuator cable not connected to
Controller
•
Blown fuse in Controller
•
Cable damaged
•
Connect Actuator cable to the
Controller
•
Fuse must be changed
•
Send Actuator for repair
2) Actuator motor runs, piston rod
does not move
•
Internal parts worn or damaged
•
Send Actuator for repair
3) Actuator cannot lift load
•
Motor worn or damaged
•
Dead Battery or Low Battery
Charge
•
Replace Actuator
•
Recharge Battery
4) Actuator piston rod only moves
inwards not outwards
•
Internal parts worn or damaged
•
Send actuator for repair
5) Pressing Handset button, Actuator
does not operate and/or no relay
noise is heard from Controller
•
Controller is defective
•
Handset is defective
•
Handset not properly plugged
into Controller
•
Emergency Stop Switch
activated
•
Replace Controller
•
Replace Handset
•
Replug Handset (use lubrication
on Handset connector)
•
Release Emergency Stop knob
6) Battery will not recharge
•
Defective Battery
•
Replace Battery
7) Power indicator (green light) on
Controller does not light up.
•
Power cord not connected
•
Fuse is blown
•
Defective Power Cable
•
Defective Controller
•
Connect to mains
•
Fuse must be changed
•
Replace Cable
•
Replace Controller
8) Actuators run slowly or does not
provide full force when using
Battery Power
•
Battery voltage low
•
Defective Actuator
•
Charge or replace Battery
•
Replace Actuator
9) Controller – one channel
functioning properly, one channel
not operating.
•
Controller is defective
•
Handset is defective
•
Replace Controller
•
Replace Handset
10) Controller does not function on
Battery, but relay clicking is
heard.
•
Battery completely dead
•
Battery defective
•
Charge Battery
•
Replace Battery
11) Legs will not operate with Boom
centered
•
Controller requires re-
initialization
•
Re-initialize Controller
12) Boom will not rotate with Legs
extended and spread
•
Controller requires re-
initialization
•
Re-initialize Controller
13) Lift Actuator is only actuator
operating
•
Controller requires re-
initialization
•
Re-initialize Controller
14) Controller beeping while operating
and plugged in to Mains
•
Controller requires re-
initialization
•
Re-initialize Controller
Содержание LIFTEM MARK IV
Страница 24: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 21 CHAPTER 5 SLING MANUAL Model 3624 Model 700...
Страница 46: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 43 FIGURE1...
Страница 47: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 44 FIGURE2 FIGURE3 FIGURE4 Handset...
Страница 48: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 45 FIGURE5...