c)  LIFT ACTUATOR does not operate 

If the LIFT ACTUATOR does not move the BOOM up or down, when depressing the HANDSET 
buttons, the cause is usually improper installation of the LIFT ACTUATOR CABLE. 
If the connector is not fully engaged, LIFT ACTUATOR will not function and a clicking sound 
will be audible in the CONTROLLER. 
This cable has a jack with a rubber O-Ring that produces substantial drag when attempting to plug 
it into the receptacle of the CONTROLLER.  This drag often prevents full engagement of the jack.  
In addition, the black or LIGHT GRAY SECURITY LOCK STRIP that “locks” in all the cable 
connectors to the CONTROLLER must be completely removed when installing the jack so it does 
not interfere with the installation. 
Before installing the LIFT ACTUATOR CABLE, apply a small amount of lubricant on the O-Ring 
of the jack.  Next, assure that the SECURITY LOCK STRIP is out of the way.  Then push the jack 
all the way home into the appropriate CONTROLLER receptacle (Refer to Figure 6, Position 5).  If 
the O-Ring of the LIFT ACTUATOR jack is not visible when viewing the CONTROLLER 
connector, the jack is properly installed. 

 

d)

 

Controller requires re-initialization 
Safe Mode 

operating anomalies indicating the need for CONTROLLER re-initialization include: 

 

LEGS

 

will not extend and/or spread, with the BOOM centered. 

  BOOM will not rotate from center to left and/or right, with the LEGS

 

extended and spread. 

  CONTROLLER beeping in the 

Safe Mode

, while operating an actuator, and the 

Battery

 voltage is 

not low or when the 115 Volt AC POWER CORD is plugged in. 

  The LIFT ACTUATOR is the only operating actuator. This actuator operates independently of the 

Safe Mode

 software. 

In the event that re-initialization is required, contact MILLENNIUM MEDICAL PRODUCTS Technical 
Support and specific guidance and directions will be provided. 

Содержание LIFTEM MARK IV

Страница 1: ...MILLENNIUM MEDICAL PRODUCTS INC PATIENT LIFTAND TRANSFER SYSTEM LIFTEM OPERATIONS MANUAL MODEL MARK IV March 2006...

Страница 2: ...Lift and Transfer Back to Bed 10 CHAPTER 3 14 USEFUL TIPS 14 1 Battery Operation 14 2 Operation by 115 Volt AC Power using Wall Socket 14 3 Operation in Case of a Dead Battery or Loss of AC Electric...

Страница 3: ...A Introduction 33 B Installation of the Scale 33 C Operate the Scale and Weigh the Patient 34 D Important Points 34 E Maintenance and Recalibration 35 F Warranty 35 CHAPTER 7 36 MAINTENANCE PROCEDURE...

Страница 4: ...and transfer a patient that might ordinarily require two or more caregivers The only contact required between patient and caregiver is the application and removal of the lifting sling As a result LIF...

Страница 5: ...he lift mechanism is rotated electrically while the patient is suspended from the MAST BOOM This requires that the receptacle which the patient is being transferred to be positioned properly with resp...

Страница 6: ...ICE OR TREATMENT PLATFORM STATION Never move LIFTEM attempt to apply CPR or use an electronic defibrillator while a patient is suspended from the MAST BOOM in a lifting sling or on a stretcher 9 Use L...

Страница 7: ...ng things in mind when you operate LIFTEM Do not rush Used properly LIFTEM will save time and help reduce injury Always follow operating instructions Always make sure caregiver using LIFTEM is trained...

Страница 8: ...n this configuration the appropriate SLING is fitted to the patient The BOOM and LIFTING BAR are positioned over the patient to allow the SLING to be attached to the ends of the LIFTING BAR There is s...

Страница 9: ...ety Precautions 1 Prior to resetting the CONTROLLER verify that the patient is calm and is not caught in any part of the Wheelchair Bed etc 2 Verify that the LEGS are fully spread and extended 3 Verif...

Страница 10: ...S can be OPEN AND EXTENDED SPREAD the LEGS to the maximum angle as shown in Figure 3 This is accomplished by pressing the LEG SPREAD OPEN button Figure 4 Both LEGS will rotate to the open position Aft...

Страница 11: ...e patient will feel more secure if the caregiver is holding onto the sling straps as the individual is lifted Safety Precautions 1 Insure that LIFTING BAR remains clear of patient while lowering and a...

Страница 12: ...the receiving device 4 Do not attempt to move LIFTEM with patient suspended from BOOM 5 Insure that LIFTING BAR remains clear of patient while lowering 6 When rotating the BOOM from left to right or...

Страница 13: ...rney or wheelchair With LIFTEM away from the patient return the BOOM to its lowest position by pressing the BOOM LIFT DOWN button and return LIFTEM to its designated storage area Charge the BATTERY by...

Страница 14: ...button EXTEND the LEGS fully by pressing the LEG EXTEND OUT button The BOOM can now be moved left or right away from the patient if necessary Position LIFTEM so one of the LEGS is between the wheels o...

Страница 15: ...rom BOOM 9 Insure that the LIFTING BAR is secured to the proper LIFTING POINT on the BOOM 10 When rotating the BOOM from left to right or reverse with a patient suspended from the lift point always jo...

Страница 16: ...ting the BOOM from left to right or reverse with a patient suspended from the LIFT POINT always jog rapidly press and release the HANDSET to reduce tendency for patient to sway during rotation d After...

Страница 17: ...olt AC Power Plug the POWER CORD into a 115 Volt AC wall socket and the system will operate as though it is on BATTERY power THIS METHOD OF POWERING LIFTEM SHOULD ONLY BE USED IN EMERGENCIES 3 Operati...

Страница 18: ...ver it is highly recommended that the following configuration be used when moving LIFTEM from one location to another The LEGS must be completely CLOSED The LEGS must be completely RETRACTED The BOOM...

Страница 19: ...m your dealer or directly from MMP 3 A detailed description of the problem and its symptoms This warranty shall not apply to any product which has been repaired or altered in any way so as in our judg...

Страница 20: ...OLLER are explained If there is a problem with the LIFT ACTUATOR not working then repeat this procedure NOTE In order to protect the CONTROLLER circuitry insure that the POWER CORD is disconnected fro...

Страница 21: ...ttery will not recharge Defective Battery Replace Battery 7 Power indicator green light on Controller does not light up Power cord not connected Fuse is blown Defective Power Cable Defective Controlle...

Страница 22: ...is the CONTROLLER channel or HANDSET button of the suspect Actuator In the case where the Actuator or the Actuator cable is bad an Audible clicking sound will be heard from the CONTROLLER when depress...

Страница 23: ...all amount of lubricant on the O Ring of the jack Next assure that the SECURITY LOCK STRIP is out of the way Then push the jack all the way home into the appropriate CONTROLLER receptacle Refer to Fig...

Страница 24: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 21 CHAPTER 5 SLING MANUAL Model 3624 Model 700...

Страница 25: ...ughly inspect the sling straps for loose or frayed threads in any stitched area While twisting loops and straps thoroughly inspect the sling straps for noticeable discoloration or fatigue Thoroughly i...

Страница 26: ...Should any of these conditions be observed immediately remove sling from service and have it inspected by the manufacturer ALL REPAIRS MUST BE PERFORMED BY THE MANUFACTURER MILLENNIUM MEDICAL PRODUCTS...

Страница 27: ...pplication Instructions for Sling Model 3624 General Although LIFTEM is designed to be operated by a single caregiver Millennium Medical Products Inc recommends that at least two caregivers be present...

Страница 28: ...1 Positioning the Sling for Patient in Bed 1 If lifting from a wheelchair apply brakes on chair Lean the patient forward enough to slide the sling with handles on the outside down behind the patient s...

Страница 29: ...ht refer to the patient s left and right Configuring the Sling for Normal Use 1 Run the right leg strap under the patient s right leg and attach to the left front hook Figure 3 Note For commode use at...

Страница 30: ...for loose or frayed threads in any stitched area While twisting loops and straps thoroughly inspect the sling straps and buckles for noticeable discoloration or fatigue Thoroughly inspect the sling s...

Страница 31: ...Warranty for Sling Model 700 Millennium Medical Products Inc warrants that all slings are free from defects in materials and workmanship for a period of six months from the date of purchase This slin...

Страница 32: ...f the openings provided on the side of the FABRIC STRETCHER SLING Two straps on each side of the FABRIC STRETCHER SLING are used to assemble the FABRIC STRETCHER SLING to the ARCH END HOOKS These stra...

Страница 33: ...until the patient sling combination is approximately level A spring loaded locking plunger on one side of the MCGL is used with a series of eleven holes in the 2 square tube that secures it in fixed...

Страница 34: ...I U M M E D I C A L P R O D U C T S I N C 31 STRETCHER SLING SYSTEM CENTER RAIL AND MOVEABLE CENTER OF GRAVITY LOCATOR ARCH Arch Mounting to Moveable Center of Gravity Locator End Hooks U Bracket Loc...

Страница 35: ...isual inspection is made Prior to first visual inspection you may want make extra copies to have on hand Retain the Data Sheet s on file to make sure visual inspections are continuous and up to date S...

Страница 36: ...t The readout is digital and provides weight in increments of 0 1 kilograms and 0 1 pounds The patient s weight will be displayed for a period of approximately 60 seconds The SCALE will provide up to...

Страница 37: ...are necessary to weigh a patient a Viewing the SCALE s digital readout panel select the appropriate readout in kg or lbs by pressing the Left Button under the front panel of the scale The units of kg...

Страница 38: ...rated accuracy requirements This accuracy is stated as 0 1 of the reading For example a reading of 200 pounds indicates the true weight is between 199 8 pounds 200 0 2 to 200 2 pounds 200 0 2 Occasion...

Страница 39: ...al The Verification Procedure consists of RECORDS REVIEW PHYSICAL CHECKS AND MEASUREMENTS OPERATIONAL TEST SAFETY SOFTWARE VERIFICATION Data Sheets are provided on pages 39 and 40 to record the result...

Страница 40: ...S fully 5 Measure CENTER to CENTER WIDTH between FRONT CASTER Mounting BOLTS 6 Position the BOOM in the CENTER and Lower the BOOM until it stops 7 Measure distance between FLOOR and 600 Pound LIFT POI...

Страница 41: ...uld move slightly and stop 8 SPREAD LEGS fully 9 ROTATE BOOM fully to right 30 10 ROTATE BOOM fully to left 30 11 RETRACT LEGS Should move slightly and stop 12 CLOSE LEGS Should move slightly and stop...

Страница 42: ...in place and LEGIBLE Yes Laminated LIFTEM Key Operation Points TAG on HANDLE BAR Yes Hanging STOP SIGN on CHEEK PLATE Yes DECAL LABEL Installation Yes Product Company Name Serial No Yes HAND CONTROLL...

Страница 43: ...ROTATE BOOM to left and right Should move slightly and stop Yes 3 Spread LEGS fully Yes 4 ROTATE BOOM to left and right Should move slightly and stop Yes 5 Close LEGS completely Yes 6 Extend LEGS ful...

Страница 44: ...of the FRONT MAST BUSHING using the lubricant specified above Push the MAST in the direction away from the legs and replace the FRONT MAST BUSHING Replace the two MAST BUSHING CLAMP PLATES and screws...

Страница 45: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 42 LIFTEM Figures MILLENNIUM MEDICAL PRODUCTS INC 790A PARKWAY BLVD BROOMALL PA 19008 PH 610 541 0170 FX 610 561 0166 www mmpliftem com...

Страница 46: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 43 FIGURE1...

Страница 47: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 44 FIGURE2 FIGURE3 FIGURE4 Handset...

Страница 48: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 45 FIGURE5...

Страница 49: ...rconnection diagram for the CONTROLLER The five 5 cables for each Actuator Battery Handset and Power Cord Function Top Plug 115V Power Battery Battery Power Cord 1 Leg Extend Actuator 2 Leg Extend Act...

Страница 50: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 47 LIFTEM SCALE FIGURE7 Kwik Lok Pin Push to Release Pin...

Страница 51: ...ld be corrected If worn out damaged or deformed parts are found on the hoist or actuator then replace those parts before operating the lift again Worn Bushing Worn Piston rod eye 2 Scratches and damag...

Страница 52: ...L P R O D U C T S I N C 49 Contact Information Millennium Medical Products Inc 790A Parkway Blvd Broomall PA 19008 Phone 610 541 0170 Toll Free 888 265 7944 Fax 610 541 0166 www mmpliftem com Info mmp...

Отзывы: