M I L L E N N I U M M E D I C A L P R O D U C T S , I N C .
40
7. Measure distance between FLOOR and 600 Pound LIFT POINT HOLE CENTER.
(36” Maximum)
[or 38.5” Maximum if
Boom Height Adjustment
has been made]
8. Raise the BOOM until it STOPS. LIFT ACTUATOR should have traveled to its
full extension.
(Yes)
9. Measure distance between FLOOR and 600 Pound LIFT POINT HOLE CENTER.
(80” Minimum)
[88.5” Minimum if Boom
Height Adjustment has
been made]
10. Measure extension of the LIFT ACTUATOR from the GROOVE near END of the
PISTON SHAFT to FRONT of CAP on LIFT ACTUATOR HOUSING.
(15.69”+/-1/16”)
11. Inspect the LIFT ACTUATOR Piston Rod Eye, Bushing and Case Back Fixture for
wear. (See Notes A & B below and LIFTEM Figure 8, Page 49.)
(None observed)
12. Inspect the inner and outer tubes of the LIFT ACTUATOR for deep scratches or
dents. (See Notes A & B below and LIFTEM Figure 8, Page 49.)
(None observed)
13. Rotate the BOOM FULLY LEFT 30°.
(30° ± 1° left)
14. Rotate the BOOM FULLY RIGHT 30°.
(30 ± 1° right)
15. Return the BOOM to the CENTER POSITION and LOWER.
(Yes)
16. RETRACT and CLOSE LEGS.
(Yes)
D. SAFETY SOFTWARE VERIFICATION (See Note C below.)
1. Insure that LEGS are RETRACTED and CLOSED.
(Yes)
2. Using HANDSET; ROTATE BOOM to left and right. Should move slightly and
stop.
(Yes)
3. Spread LEGS fully.
(Yes)
4. ROTATE BOOM to left and right. Should move slightly and stop.
(Yes)
5. Close LEGS completely.
(Yes)
6. Extend LEGS fully.
(Yes)
7. ROTATE BOOM to left and right. Should move slightly and stop.
(Yes)
8. SPREAD LEGS fully.
(Yes)
9. ROTATE BOOM fully to right (30°).
(Yes)
10. ROTATE BOOM fully to left (30°).
(Yes)
11. RETRACT LEGS. Should move slightly and stop.
(Yes)
12. CLOSE LEGS. Should move slightly and stop.
(Yes)
13. Return BOOM to center position.
(Yes)
14. Retract and close LEGS.
(Yes)
NOTES: A.
If deep inner or outer tube scratches, dents or elongated/worn bushings/holes are observed, the parts are to
be removed and replaced.
B.
If any unusual or abnormal sounds such as ticking, grinding or screeching occur when operating in the
loaded or unloaded mode the LIFT ACTUATOR is to be removed and replaced.
C.
Failure of any item in Section D indicates that CONTROLLER SAFETY SOFTWARE must be
RE-INITIALIZED. Record any noises or abnormalities observed during the performance of this procedure.
15.
Record Comments:
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
_____________________________________________________________
Содержание LIFTEM MARK IV
Страница 24: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 21 CHAPTER 5 SLING MANUAL Model 3624 Model 700...
Страница 46: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 43 FIGURE1...
Страница 47: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 44 FIGURE2 FIGURE3 FIGURE4 Handset...
Страница 48: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 45 FIGURE5...