f) Safety Software Verification:
The purpose of the Safety Software Verification Test is to verify that the Safety Software features
built into the Controller are fully functional. Record all results on the Data Sheet.
1)
Insure that LEGS are RETRACTED and CLOSED.
2)
Using HANDSET; ROTATE BOOM to left and right. BOOM should move slightly and stop.
3)
Spread LEGS fully.
4)
ROTATE BOOM to left and right. Should move slightly and stop.
5)
Close LEGS completely.
6)
Extend LEGS fully.
7)
ROTATE BOOM to left and right. Should move slightly and stop.
8)
SPREAD LEGS fully.
9)
ROTATE BOOM fully to right (30°).
10)
ROTATE BOOM fully to left (30°).
11)
RETRACT LEGS. Should move slightly and stop.
12)
CLOSE LEGS. Should move slightly and stop.
13)
Return BOOM to center position.
14)
Retract and close LEGS.
NOTES: A. If deep inner or outer tube scratches, dents or elongated/worn bushings/holes
are observed, the parts are to be removed and replaced.
B. If any unusual or abnormal sounds such as ticking, grinding or screeching occur
when operating in the loaded or unloaded mode the LIFT ACTUATOR is to be
removed and replaced.
C. Failure of any item in Section F indicates that CONTROLLER SAFETY
SOFTWARE must be RE-INITIALIZED. Record any noises or abnormalities
observed during the performance of this procedure.
Содержание LIFTEM MARK IV
Страница 24: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 21 CHAPTER 5 SLING MANUAL Model 3624 Model 700...
Страница 46: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 43 FIGURE1...
Страница 47: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 44 FIGURE2 FIGURE3 FIGURE4 Handset...
Страница 48: ...M I L L E N N I U M M E D I C A L P R O D U C T S I N C 45 FIGURE5...