Disassembly Instructions
Repair Information
4 - 16
0070-10-0591-01
Trio™ Service Manual
4.5.10
Removal of NIBP/IBP PCB Mounting Plate
1.
Remove two (2) 3 x 6 mm Phillips panhead screws at upper right side center and upper
left side center of PCB chassis assembly.
2.
Remove one (1) 3 x 6 mm Phillips panhead screw at lower left of NIBP/IBP pcb
mounting plate.
3.
If IBP PCB is installed, remove cable assemblies from PD1 and P2 of IBP PCB.
4.
Remove cable assembly from P1 of NIBP PCB. Remove polyurethane hose from Rectus
fitting on parameter panel.
5.
Lift NIBP/IBP PCB Mounting Plate up and out to remove. This will allow access to ECG/
RESP/TEMP PCB and SpO
2
PCB.
4.5.11
Removal of Handle
1.
Remove two (2) 4 x 14 mm Phillips panhead self tapping screw at top inside front of
rear housing.
2.
Push handle and handle shoes toward rear of rear housing until they are removed from
housing.
3.
Use an awl or similar tool to push the pins through the shoes releasing the handle.
Содержание Trio
Страница 1: ...Service Manual 0070 01 0591 02_revB_Trio srv color indd 1 3 10 10 5 36 48 PM ...
Страница 2: ...Service Manual ...
Страница 7: ...Introduction Warning iv 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 55: ...Introduction Repair Information 4 2 0070 10 0591 01 Trio Service Monitor This page intentionally left blank ...
Страница 58: ...Trio Service Manual 0070 10 0591 01 4 5 Repair Information Introduction FIGURE 4 3 CPU PCB Connector Reference ...
Страница 61: ...Introduction Repair Information 4 8 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 81: ...0070 10 0591 01 Revision G July 17 2010 ...