Calibration and System Checks
Calibration/Maintenance
2 - 6
0070-10-0591-01
Trio™ Service Manual
NIBP CAL
Connect a test chamber (P/N 0138-00-0001-03) and a calibrated digital mercury
manometer with bulb via T fitting to the NIBP quick connect fitting of the
Trio
Monitor
(Figure 2-5.)
1.
From the
MAINTENANCE
menu select
NIBP
CAL
.
The selected tile will now read
STOP CAL
. The user may select this tile at any time to stop the test.
2.
Using the bulb, inflate the pneumatic system so that the digital mercury manometer reads
0, 50 and 200 mmHg in turn. The difference between the indicated pressure of the
digital mercury manometer and the indicated pressure in the
NIBP
parameter tile of the
Trio
Monitor should not exceed 3 mmHg. If it exceeds 3 mmHg, please contact a
service technician.
FIGURE 2-5
Diagram of NIBP Calibration
DIGITAL OR
MERCURY
MANOMETER
500 ml.
TEST CHAMBER
P/N 0138-00-0001-03
BULB
hose
TRIO MONITOR
NIBP
Содержание Trio
Страница 1: ...Service Manual 0070 01 0591 02_revB_Trio srv color indd 1 3 10 10 5 36 48 PM ...
Страница 2: ...Service Manual ...
Страница 7: ...Introduction Warning iv 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 55: ...Introduction Repair Information 4 2 0070 10 0591 01 Trio Service Monitor This page intentionally left blank ...
Страница 58: ...Trio Service Manual 0070 10 0591 01 4 5 Repair Information Introduction FIGURE 4 3 CPU PCB Connector Reference ...
Страница 61: ...Introduction Repair Information 4 8 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 81: ...0070 10 0591 01 Revision G July 17 2010 ...