Testing Each Parameter
Calibration/Maintenance
2 - 12
0070-10-0591-01
Trio™ Service Manual
2.5
Testing Each Parameter
Testing each parameter annually ensures the accuracy of the
Trio
Monitor. The following
tests should be used for verification of operation only. The following tests should not be
considered calibration procedures.
2.5.1
ECG and RESP
2.5.1.1
Test Equipment
• Patient Simulator
2.5.1.2
Test Procedures
1.
Use a 5L patient cable and lead wires to connect a calibrated patient simulator to the
ECG connector of
Trio
.
2.
Set ECG lead to lead II and ECG size to x1.
3.
Check if ECG waveforms and RESP waveforms are displayed.
4.
Set up the parameters of the simulator as follows:
HR=30 (gain × 4)
RR=15
5.
Check if the displayed ECG and RESP waveforms, HR and RR values are correct.
6.
Change the simulator configuration:
HR=240
RR=120
7.
Check if the displayed ECG and RESP waveforms, HR and RR values are consistent with
the parameters set up on the simulator.
8.
Remove ECG lead. In this condition, the
Trio
should immediately alarm and display
ECG LEAD OFF
at the top right side of the Message Area.
2.5.2
NIBP
2.5.2.1
Test Equipment
• NIBP Simulator
• NIBP Dummy Cuff/Chamber
Содержание Trio
Страница 1: ...Service Manual 0070 01 0591 02_revB_Trio srv color indd 1 3 10 10 5 36 48 PM ...
Страница 2: ...Service Manual ...
Страница 7: ...Introduction Warning iv 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 55: ...Introduction Repair Information 4 2 0070 10 0591 01 Trio Service Monitor This page intentionally left blank ...
Страница 58: ...Trio Service Manual 0070 10 0591 01 4 5 Repair Information Introduction FIGURE 4 3 CPU PCB Connector Reference ...
Страница 61: ...Introduction Repair Information 4 8 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 81: ...0070 10 0591 01 Revision G July 17 2010 ...