Trio™ Service Manual
0070-10-0591-01
2 - 9
Calibration/Maintenance
Calibration and System Checks
Error Log:
This menu will display any system errors which may have occurred during
normal operation of unit.
SCROLL
or
this information by selecting
the respective tiles
FIGURE 2-9
Error Log Menu*
Factory:
This menu allows the unit to be preconfigured at the factory. This menu is
enabled with an access code, 762718, and should only be accessed as
necessary by Customer Service Personnel or by trained Bio-medical
technicians after replacing defective modules as a result of unit repair.
The Factory menu allows for configuration of:
Serial Port:
This menu allows the user to configure the DIAP baud rate of the serial
output port. The choices are 9600 bps or 19200 bps.
• Language (English, Spanish*, French*, German*, Italian*, Chinese*).
• Module Options (ECG, Resp, SpO
2
, 1ch.Temp, 2ch.Temp*, NIBP, IBP,
Recorder).
Selections are enabled or disabled by checking or unchecking the box
which corresponds to each item in the menu.
NOTE:
IBP and 2ch. Temp should not be enabled
simultaneously.
*
Not available for domestic units.
Содержание Trio
Страница 1: ...Service Manual 0070 01 0591 02_revB_Trio srv color indd 1 3 10 10 5 36 48 PM ...
Страница 2: ...Service Manual ...
Страница 7: ...Introduction Warning iv 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 55: ...Introduction Repair Information 4 2 0070 10 0591 01 Trio Service Monitor This page intentionally left blank ...
Страница 58: ...Trio Service Manual 0070 10 0591 01 4 5 Repair Information Introduction FIGURE 4 3 CPU PCB Connector Reference ...
Страница 61: ...Introduction Repair Information 4 8 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 81: ...0070 10 0591 01 Revision G July 17 2010 ...