Trio™ Service Manual
0070-10-0591-01
5 - 5
Appendix
System Alarm Prompts
LOOSE CUFF
The NIBP cuff is not connected
correctly
Re-connect the NIBP cuff.
AIR LEAK
The NIBP cuff is not connected
correctly or there are leaks in
the airway
Check the connection of each
part or replace cuff and hose.
If message does not clear
contact Customer Service.
OVER PRESSURE
NIBP is in an overpressure
condition.
Check for occlusion in cuff
and hose. Measure again. If
message does not clear
contact Customer Service.
SIGNAL SATURATED
Problem occurs when
measuring NIBP. The system
cannot perform measurement,
analysis or calculation.
Check the connection of the
cuff and hose. Check
patient's condition. Measure
again, if the failure persists,
contact the manufacturer for
repair.
TIME OUT
Problem occurs when
measuring NIBP. The system
cannot perform measurement,
analysis or calculation.
Check the connection of the
cuff and hose. Check
patient's condition. Measure
again, if the failure persists,
contact the manufacturer for
repair.
CUFF TYPE ERR
NIBP cuff may be wrong size
Check if the patient size is set
correctly. Replace existing
cuff with correct size cuff. If
the failure persists, contact
the manufacturer for repair.
PNEUMATIC LEAK
NIBP airway has leaks
Check the connection of cuff
and hose or replace cuff and
hose. If the failure persists,
contact the manufacturer for
repair.
MEASURE FAIL
Problem occurs when
measuring NIBP. The system
cannot perform measurement,
analysis or calculation.
Check the connection of the
cuff and hose. Check
patient's condition. Measure
again, if the failure persists,
contact the manufacturer for
repair.
NIBP SYSTEM FAILURE
Problem occurs when
measuring NIBP. The system
cannot perform measurement,
analysis or calculation.
Check the connection of the
cuff and hose. Check
patient's condition. Measure
again, if the failure persists,
contact the manufacturer for
repair.
PROBLEM/MESSAGE
REASON
SOLUTION
*
XX represents all the parameter modules in the system such as ECG, NIBP, SpO
2
, IBP module, etc.
Содержание Trio
Страница 1: ...Service Manual 0070 01 0591 02_revB_Trio srv color indd 1 3 10 10 5 36 48 PM ...
Страница 2: ...Service Manual ...
Страница 7: ...Introduction Warning iv 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 55: ...Introduction Repair Information 4 2 0070 10 0591 01 Trio Service Monitor This page intentionally left blank ...
Страница 58: ...Trio Service Manual 0070 10 0591 01 4 5 Repair Information Introduction FIGURE 4 3 CPU PCB Connector Reference ...
Страница 61: ...Introduction Repair Information 4 8 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 81: ...0070 10 0591 01 Revision G July 17 2010 ...