Calibration and System Checks
Calibration/Maintenance
2 - 4
0070-10-0591-01
Trio™ Service Manual
IBP Pressure Cal
The purpose of this calibration is to ensure that the system gives you accurate measurements.
FIGURE 2-3
IBP Calibration
WARNING: Never perform this procedure while patient is being
monitored.
1.
Connect a Pressure Transducer Interface Cable from a Pressure Transducer
(Figure 2-3) to the IBP socket on the right side panel of the
Trio
2.
Open 3-Way Stopcock to atmosphere.
3.
Perform zeroing procedure by selecting the
IBP
parameter menu, then selecting
IBP
ZERO
tile.
4.
From
IBP ZERO
menu select
IBP ZERO
tile to zero the blood pressure channel.
5.
Connect a Sphygmomanometer or a digital Manometer w/bulb to the Pressure
Transducer via a T fitting.
6.
Close the 3-way Stopcock.
7.
Manually pump the Manometer to a static value of 100 mmHg and close valve on bulb.
8.
From the
MAINTENANCE
menu, select
IBP PRESSURE CAL
.
9.
From
IBP PRESSURE CAL
menu, select the tile adjacent to
CAL VALUE
and set to 100
mmHg.
Содержание Trio
Страница 1: ...Service Manual 0070 01 0591 02_revB_Trio srv color indd 1 3 10 10 5 36 48 PM ...
Страница 2: ...Service Manual ...
Страница 7: ...Introduction Warning iv 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 55: ...Introduction Repair Information 4 2 0070 10 0591 01 Trio Service Monitor This page intentionally left blank ...
Страница 58: ...Trio Service Manual 0070 10 0591 01 4 5 Repair Information Introduction FIGURE 4 3 CPU PCB Connector Reference ...
Страница 61: ...Introduction Repair Information 4 8 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 81: ...0070 10 0591 01 Revision G July 17 2010 ...