Trio™ Service Manual
0070-10-0591-01
2 - 3
Calibration/Maintenance
Calibration and System Checks
FIGURE 2-2
Maintenance Menu (below S/N MC15000-XX)
2.4.2.1
Calibrations
Calibrations should be performed at least once a year. Calibrations should also be
performed after any preventive maintenance or repair of the
Trio
Monitor.
From the
MAINTENANCE
menu make the following selections to perform each specific
calibration or check.