Trio™ Service Manual
0070-10-0591-01
5 - 1
5.0
Appendix
5.1
System Alarm Prompts
PROBLEM/MESSAGE
REASON
SOLUTION
BATTERY VOLTAGE TOO LOW
When battery voltage is too
low, the monitor will
automatically shut down within
5 minutes
Use AC power supply
ECG WEAK SIGNAL
The patient's ECG signal is too
weak
Check if the electrodes and
lead wires are connected
correctly. Check the current
condition of the patient.
NO PULSE
The patient's pulse signal is too
weak or non-existent
Check the connection of the
sensor. Check the current
condition of the patient.
PNP
The pacemaker is not paced
Check the connection of the
pacemaker. Check the
connection of electrodes and
lead wires. Check the current
condition of the patient.
PNC
No pacemaker signal is
captured
Check the connection of the
pacemaker. Check the
connection of electrodes and
lead wires. Check the current
condition of the patient.
ECG LEAD OFF
ECG lead is not connected
correctly
Check the connection of ECG
lead wires
ECG V LEAD OFF
The V lead wire of ECG is not
connected correctly
Check the connection of V
lead wire
ECG LL LEAD OFF
The LL lead wire of ECG is not
connected correctly
Check the connection of LL
lead wire
ECG LA LEAD OFF
The LA lead wire of ECG is not
connected correctly
Check the connection of LA
lead wire
*
XX represents all the parameter modules in the system such as ECG, NIBP, SpO
2
, IBP module, etc.
Содержание Trio
Страница 1: ...Service Manual 0070 01 0591 02_revB_Trio srv color indd 1 3 10 10 5 36 48 PM ...
Страница 2: ...Service Manual ...
Страница 7: ...Introduction Warning iv 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 55: ...Introduction Repair Information 4 2 0070 10 0591 01 Trio Service Monitor This page intentionally left blank ...
Страница 58: ...Trio Service Manual 0070 10 0591 01 4 5 Repair Information Introduction FIGURE 4 3 CPU PCB Connector Reference ...
Страница 61: ...Introduction Repair Information 4 8 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 81: ...0070 10 0591 01 Revision G July 17 2010 ...