Calibration and System Checks
Calibration/Maintenance
2 - 2
0070-10-0591-01
Trio™ Service Manual
2.4
Calibration and System Checks
2.4.1
Device Appearance and Installation Checks
Inspect the
Trio
Monitor to ensure that:
• The outer housing of the device is clean and has no scratches or cracks
• When shaking the device, there are no loose components
• All keys are smooth and free for operation
• Labels are complete, sufficient and accurate
• Standard configuration is complete and all connectors are installed securely
2.4.2
Maintenance Menu
The
MAINTENANCE
menu provides access to all user operable calibration checks. It also
provides access to certain technical information and settings. To access the
MAINTENANCE
menu, turn the unit on, rotate the Navigator™ knob to highlight the
MENU
icon at the bottom of the screen and push the knob to select the
SYSTEM MENU
.
From the
SYSTEM MENU
, select
MAINTENANCE
.
FIGURE 2-1
Maintenance Menu DIAP units (S/N MC15000-XX and above)
Содержание Trio
Страница 1: ...Service Manual 0070 01 0591 02_revB_Trio srv color indd 1 3 10 10 5 36 48 PM ...
Страница 2: ...Service Manual ...
Страница 7: ...Introduction Warning iv 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 55: ...Introduction Repair Information 4 2 0070 10 0591 01 Trio Service Monitor This page intentionally left blank ...
Страница 58: ...Trio Service Manual 0070 10 0591 01 4 5 Repair Information Introduction FIGURE 4 3 CPU PCB Connector Reference ...
Страница 61: ...Introduction Repair Information 4 8 0070 10 0591 01 Trio Service Manual This page intentionally left blank ...
Страница 81: ...0070 10 0591 01 Revision G July 17 2010 ...