PRO
2
XY OXYGEN CONCENTRATOR
519211-22
3
EN
MEANING OF SYMBOLS USED IN THIS MANUAL
Symbols are used in this manual to help you understand them and to indicate particular points which you need to act upon or
consider. Those symbols are:
Reading symbol:
to refer to the operator manual / instructions manual.
Danger symbol
: This symbol represents the DIN 4844 symbol (hazard warning); it warns of a hazard which could result in
death or injury, along with damage to the machine. This symbol must be observed by all personnel working on this machine.
Regulations on safety in the workplace must be observed.
Environmental protection symbol:
This symbol serves as a reminder to sort waste during maintenance operations, to store
it in a safe place, and to dispose of it with due care for the environment.
Dustbin on wheels barred of a cross symbol:
This symbol indicates that the equipment should not be thrown with domestic
waste and is subject to selective collection.
Symbol prohibition to lubricate
: This pictogram indicates prohibition to use grease or lubricant.
Non-ionizing radiation hazard symbol
:
This pictogram informs of a risk or a danger related to non-ionizing radiations.
Medical Device
: The product is a medical device.
Additional pictograms on the oxygen concentrator
Mandatory:
use noise protection
(ISO 7010-M003)
Mandatory:
read operator
instructions
(ISO 7000-M002)
Warning:
risk or danger of
possible presence of
combustion agent
Warning:
risk of electric shock
(ISO 7010-W012).
Warning:
do not touch hot
surfaces
(danger of burns)
(ISO 7010-W017).
It is formally forbidden to take off the adhesive labels being on this machine.
The deteriorated adhesives or falling apart will have to be replaced.
General remark:
If used in accordance with their intended purpose, units bearing the CE mark will satisfy the essential requirements of directive
93/42/EEC, relative to medical devices and, where necessary, to essential requirements of the directive 2014/68/EU, relative to
pressure equipment.
Those units,
when installed in accordance with these instructions
, comply with the provisions of currently applicable European
Directives and harmonised standards.
However, your attention is drawn to the fact that a unit with only one compressor or only one vacuum pump cannot be used without
an emergency backup device: gas backup or other fixed or mobile unit.
For devices intended to be fixed and connected to a medical distribution system, the CE marking provided by MIL'S as well as the CE
conformity declaration of the product covered by the start-up and maintenance instructions becomes invalid for any relocation to
another site/health establishment.
The device may only be relocated to another site if it has been reconditioned under the manufacturer's responsibility to comply with
general safety and performance requirements, please contact us for more information.