ELECTRICAL CONNECTIONS
Work on the unit should only take place with the electricity supply off.
Any connection work should be done only by the exploitaing.
Class I device: to avoid any risk of electric shock, this device must only be connected to a power supply network
equipped with a protective earth.
To avoid any risk of electric connections having become loose from vibration during transport, we advise you to check
the tightness of all electric connections before starting up.
Protect all power leads located upstream by an over-current protection device in compliance with the legislation and
regulations in force in your country.
To avoid any risk of electric shock, this device must be only connected to a supply network equipped with a protection
ground.
8.1
Connection HOSPITAIR PACK S / PROCOM 2 Concentrator
400V three-phase + Ground
supply for each screw compressor: cable cross-section according to current and length.
Connection between the compressor and the concentrator is done via a connector.
All items such as air production units and O
2
concentrator units should be connected to an electric supply that has a
back-up generating unit.
5 bar Version
Power supply: .......................... 230V 1-phase + G
Power: ..................................... 150 W
Cable cross-section: ................. 0.75 mm²
12 bar Version
Power supply: .......................... 400V + N + G
Model
OX3B / OX5B
OX7B/OX9B/OX13B/OX16B
OX26B / OX32B
Power
700 W
1650 W
2350 W
Cable cross-section
1.5 mm²
1.5 mm²
2.5 mm²