XY OXYGEN CONCENTRATOR
28
519211-22
EN
To change value of parameters, press the
button. Increase or decrease the value by the
or
button. To valid, press
the
button.
If the
button is not activated for 2 minutes, the fourth line will automatically turn blank.
GAS ANALYSERS (OPTIONAL)
13.1
Regulation
The mandatory ISO 7396-1: 2016 standard requires the monitoring of:
➢
Oxygen rate produced by each concentrator at its output:
O2
sensor
➢
Oxygen rate produced by all the lines, i.e. a single sensor for the installation of one or more lines:
O2r
sensor, connected to
concentrator A.
➢
Rate of carbon monoxide in the gas produced by all the lines:
CO
probe.
➢
Ambient oxygen level in the room containing at least one concentrator:
ambient O2
probe, with an audible and visual alarm
outside the room.
These monitoring and alarm systems are offered as an option by MIL'S.
13.2
Oxygen concentrator and network sensors with 4-20mA output
Paramagnetic sensor controlled in temperature at 55°C, measures oxygen concentration. It is equipped with a barometric
compensator to avoid temperature and atmospheric pressure variations.
13.3
Other oxygen concentrator sensor with 0-1V output
The measurement of oxygen concentrator level can be carried out with a paramagnetic sensor with 0-1V output, with a barometric
compensator to avoid atmospheric pressure variations.
2
O
1067 mbar
Paramagnetic
sensor
controlled at 50°C
Barometric
compensator
Flow rate setting
at 100 ml / min (cc/min)
Regulator at 200 mbar
at 5 bar
C2O