PRO
2
XY OXYGEN CONCENTRATOR
10
519211-22
EN
3.3.3
Electric Low-Pressure Booster
When the booster is started, the independent fan starts.
A temperature sensor constantly checks booster outlet
temperature.
If temperature is above the upper limit, the compressor
is stopped.
BOOSTER Type
Oxygen concentrator
Upper limit temperature
HG26
OX3B
–
OX5B
130°C
HG50
OX7B
–
OX9B
–
OX13B
–
OX16B
130°C
HG75
OX26B
–
OX32B
130°C
When the booster is stopped, the independent fan also stops.
In the event of an electric power failure or an overheating of the booster, a red indicator lights up on the concentrator.
For electric power failure, once the problem has been solved, the circuit breaker must be reset.
For overheating, low-pressure booster is stopped then starts again when the temperature is below the upper limit temperature.
Before the first start-up, ensure proper rotation direction of the compressor fan.
3.4
Description of remainder
CpO
2
:
sensor upstream booster to control booster entry pressure. If this pressure is less than 2.7 bar (default value),
booster does not start.
Electric Booster:
electric booster to compress O
2
from a given pressure to a higher pressure.
Storage:
high pressure O
2
storage receiver. Pmax: 15 bar
C1O:
pressure sensor controlling O
2
pressure at booster exit and thus in storage receiver.
Release line:
is equipped of
•
a
PAR
ticle filter to deliver gas without particles,
•
a regulator
(D4)
to control network pressure.
C2O:
network sensor to control O2 network pressure with upper and lower alarm systems.
DIMENSIONS
4.1
Dimensions PRO
2
XY OX3
➔
OX16 oxygen concentrators
PRO
2
XY
oxygen
concentrator
OX3
OX3B
OX 5
OX 5 B
OX 7
OX7B
OX9
OX9B
OX13
OX16
OX13B
OX16B
A
1210
1250
1490
1590
B
1000
1000
1000
1010
C
2046
2180
2230
2295
D
670
670
670
660
E
960
960
960
870
Dia. Air outlet
3/4"
3/4"
1"
1"
Dia. O
2
outlet
1/2"
1/2"
1/2"
1/2"
A
C
E
B
D
Entrée
d'AIR
Sortie
O
2
outlet
air
inlet