PRO
2
XY OXYGEN CONCENTRATOR
30
519211-22
EN
Once calibration finished, don't forget to turn the handle of the 3-way valve on
produced O
2
way.
Calibration periodicity:
•
At the first start-up,
•
At the end of one month of operation,
•
Then every 6 months.
13.5
Reference gas value:
From 2009 to February 2017, the O2 probes ranged from 80% (4 mA) to 100% (20 mA).
From February 2017, the range has increased from 0% (4 mA) to 100% (20 mA).
The other oxygen concentrator sensor ranged from 0% (0V) to 100% (1V).
Gas ZERO:
•
For a probe from 2009 to 2017, we recommend a reference value between 80% and 82%. This requires a special mixing
bottle.
•
For a probe starting in 2017, we recommend a reference value of 0% from a nitrogen cylinder, or 21.9% (ambient air value,
recoverable on compressed air) for example.
Gas SPAN:
•
In both cases, we recommend using a reference value close to 100%, from a bottle of medical O
2
.
CLEANING AND MAINTENANCE
See the manual of maintenance delivered with this instruction's manual.
IMPORTANT :
For this operation, use all necessary protective measures (gloves, mask, overalls etc.) then destroy contamined
consumables (antibacterial cartridges, condensates etc.) in according with that country’s applicable laws (regulation
incinerator).
Maintenance is to be entrusted to trained and approved personnel by the exploiting, with all necessary knowledge and
acquainted with best practice.
Not to carry out maintenance operations when the device is in use.
If the spare parts used during maintenance operations are not corresponding to the manufacturers' requirements, the
manufacturer declines all responsibility in the event of an incident.
The modification of the central units is forbidden.
Modification of the device is forbidden without the approval of the manufacturer.
If the concentrator is modified following manufacturer agreement , a control and a suitable test must be realized to
make sure that the concentrator is always safe to use.
The original parts of the concentrator are all degreased. Take care not to soil them with greasy substance. All
maintenance operations on oxygen must be realized with an extreme cleanliness.
The parts must respect the cleanliness requirements defined in the standard ISO 15001.
Flowmeter
100 cc/min (=100 ml/min)