Page 54
Measure
optoCONTROL 2600
Response:
Byte 1
Byte 2
Byte 3
Byte 4
hex
Remark
„0“
„D“
„C“
„1“
0x3143444F
0x28
0xA0
0x03
0x00
0x0003A028
Without error
0x00
0x00
0x00
0x00
0x00000000
Error code
Avoid error!
> Data are not accepted!
Possible errors:
0x00000004
Too much data received
0x0000000A
Error on writing to the RAM
0x0000000B
Incorrect data transmitted, see
Valid values
6.6.3.14 Save Measurement Program Data
Name
SAVE MPR RAM TO FLASH
Description:
With this command the currently valid measurement program data
is written from the main memory to the flash memory.
Command:
Byte 1
Byte 2
Byte 3
Byte 4
hex
„+“
„+“
„+“
0x0D
0x0D2B2B2B
„0“
„D“
„C“
„1“
0x3143444F
0x2A
0x20
0x00
0x00
0x0000202A
Response:
Byte 1
Byte 2
Byte 3
Byte 4
hex
Remark
„0“
„D“
„C“
„1“
0x3143444F
0x2A
0xA0
0x03
0x00
0x0003A02A
Without error
0x00
0x00
0x00
0x00
0x00000000
Error code
Possible errors:
0x00000006
Flash - access error
6.6.3.15 Resetting the Output Values in the Trigger Measurement Mode
Name
TRIGGERMODE RESET
Description:
Resets the output values in the Trigger measurement mode
Command:
Byte 1
Byte 2
Byte 3
Byte 4
hex
„+“
„+“
„+“
0x0D
0x0D2B2B2B
„0“
„D“
„C“
„1“
0x3143444F
0x2B
0x20
0x00
0x00
0x0000202B
i
This command corresponds to the external reset input in the Trigger measurement
mode.
Response:
Byte 1
Byte 2
Byte 3
Byte 4
hex
Remark
„0“
„D“
„C“
„1“
0x3143444F
0x2B
0xA0
0x03
0x00
0x0003A02B
Without error
0x00
0x00
0x00
0x00
0x00000000
Error code
NOTICE