Page 10
Measure
optoCONTROL 2600
3.3.2
Rear View of the Controller
Receiver (12-pole)
Inputs and outputs (25-pole)
Light source (5-pole)
Supply voltage (3-pole)
Fig. 5 Connection on the rear side of the controller
3.4
Operating Modes
The following operating modes are selectable via a menu-assisted selection (measure-
ment program
):
-
Position of an edge (bright/dark or dark/bright)
-
Diameter of a target
-
Gap between two targets
-
Distance between two selectable edges (segment)
-
Serial measurement of up to four freely selectable segments (multi-segment) via the
digital output (e.g. segments 1 - 4 and 2 - 3)
i
Factory setting:
Position edge bright - dark
Re
cei
ver
Recei
ver
Recei
ver
Recei
ver
1
2
3
4
Fig. 6 Methods of operation
For each measurement program 2 limits and 2 warning levels can be programmed. For
the
Multi-segment
program only 2 limits per segment 1 and segment 2 can be pro-
grammed.
Potentially measured segments 3 and 4 are not monitored.
Application-specific measurement programs can also be generated by menu.