Remote programming
Function Generators
XIII -
53
Remote programming (programmable version only) (contd.)
The programming instructions respect the IEEE 488-2 standard and the
SCPI (
Standard Commands for Programmable Instruments
) protocol.
The user has the possibility of having complete remote control of the device.
For more information please consult the programming guide.
Communication
interfaces
Connecting the generator to a PC is done using either :
•
an A/B type USB cable via a USB to UART converter,
•
via ETHERNET Warning ! To use the ETHERNET link, the USB
cable must be disconnected.
USB
If the CP210x driver is correctly installed on the PC the USB peripheral will
be recognised and a new COM port will appear in the PC’s system settings
(see the programming guide to install).
The new COM port is configured as follows:
- speed: 19200 bauds
- data bits: 8
- parity: none
- stop bit: 1
- protocol: hardware (RTS / CTS)
ETHERNET
Once the IP address has been programmed through the GX320E-Admin
application, the GX320E can be accessed via this address.
Connection
via USB
USB Connector
USB
Port
USB
Interface
cable