General Commands
IV - 24
Function Generators
General Commands (contd.)
In the event of an error the automatic calibration stops at the step in error, it
then switches to manual mode.
The display shows:
In the event of repeated errors contact your CHAUVIN-ARNOUX representative
(see p. 5).
Saving settings
(GX 320)
(GX 305/310)
Pressing saves the calibration.
The
display is cleared once the backup is made. It reappears if the
calibration is changed.
Exit from
Calibration mode
Exit this mode using this key.
The key becomes:
.
To save settings a data backup should be made (see above) before
exiting the mode, otherwise the settings are lost and the previous
settings are re-loaded at start-up.