General commands
IV - 28
Function Generators
General Commands (contd.)
Saving a configuration
(GX 320)
The GX 320 can save and reload user configurations.
A total of up to 15 files can be saved.
This backup is permanent (the data is saved even if the instrument is powered
down).
Enters the configuration management mode.
is displayed on the screen with the current file number:
if file 3 is empty;
if file 3 already contains a configuration
the data (other than frequency) is displayed on the screen.
Pressing another key than
or
exits the mode without saving.
Selects files from SEt.01 to SEt.15. The screen is updated with the data from
the selected file.
Pressing the key again saves the current configuration in the selected file.
The display returns to its pre-backup status and the
display is cleared.
When saving the content of the selected file is overwritten by the content
of the current configuration without any warning messages.