BURST Function
XII - 52
Function Generators
BURST Pulse burst function (contd.)
Setting the number of
pulses Num
The pulse number value (Num) can be limited in INTernal source by the
value of the entered period (Time).
In both cases (INTernal or EXTernal), the Num
min
value is set in order to
avoid having a window of less than 2 µs (see above).
or
Display of the number of pulses Num and assignment of the adjustment to
the wheel.
The key lights:
.
Pressed successively selects the digit to which the wheel increments will be
applied.
Value adjustment.
Pressed successively selects the digit to which the wheel increments will be
applied.
1s
Using an INT source pressing successively for more than 1 second switches
from Num to Time and vice versa, otherwise selects Num setting.
Setting the generation
period using an INTernal
source
1s or
1s
Displays the Time and assigns the wheel to adjustment.
The key lights:
.
1s
Pressed successively switches from Num to Time.
Pressed successively selects the digit to which the wheel increments are
applied.
Value adjustment.
1s
Switches from Num to Time and vice versa.
Manual triggering using
EXTernal source
Pressing this key triggers the generation of a pulse burst.
Other settings
See the CONT function.