Generation of basic periodic signals
V - 36
Function Generators
Generation of basic “CONT” periodic signals (contd.)
Setting the offset and
DC level
or
Assignment of offset adjustment to the wheel. The
key lights.
Value adjustment.
The variation field is from -10 V to +10 V maximum in an open circuit.
1s
Forces the offset value to 0.
The sum of continuous v alternating voltage cannot be
>>>>
± 10 V.
Setting signal logical
levels
This function is only available if the “LOGIC” waveform has been selected.
or
Assignment of the logic signal low level to the wheel.
The
key lights.
The "Adj.LO" message displays instead of the frequency value:
By pressing several times the high or low level is selected, "Adj.HI" is displayed
for high level adjustment:
Adjustment of the selected value.
The field of variation for these levels is from -10 V to +10 V by 100 mV intervals.
The high level is always greater than or equal to the low level.