Menu mode
5. Set the calibration value to 0.0 K and ac-
cept the setting by pressing the confirma-
tion key.
C
K
0,2
-
C
K
C
K
0,2
-
e
Cal1
Cal2
Cal3
40.0
5.0
30.0
0,0
6. Position the sensor of a calibrated refer-
ence instrument centrally in the appli-
ance's working chamber.
7. Close the door and, in manual mode,
adjust the set temperature to 30 °C.
21
.4°C
TEMP
30
.0
°C
Set
8. Wait until the appliance reaches the set
temperature and displays 30 °C. The
reference instrument for example displays
31.6 °C.
30
.0°C
TEMP
Set 30
.0°C
31.6
°C
9. In the SETUP, adjust the calibration value
Cal2 to +1.6 K (actual value measured
minus setpoint temperature) and save the
setting by pressing the confirmation key.
C
K
0,2
-
C
K
C
K
0,2
-
e
Cal1
Cal2
Cal3
40.0
5.0
30.0
1,6
+
10. After the calibration procedure, the
temperature measured by the reference
instrument should now also be 30 °C.
30
.0°C
TEMP
Set 30
.0°C
30,0
°C
With Cal1, a calibration temperature below Cal2 can be programmed accordingly, and with
Cal3, a temperature above. The minimum difference between the Cal values is 10 K.
If all calibration values are set to 0.0 K, the factory calibration settings are restored.