D33464 | Date 09/2018
33
Operation and control
Adjusting temperature monitoring
Press the activation key to the left of the
ALARM
display. The temperature monitoring
setting is automatically activated ( ).
min
000°C
ALARM
max
000
°C
auto off
99K
-
+
ALARM
max
40
.0°C
auto
0.0
K
min
15
.0°C
+-
Save the selection by pressing the confirma-
tion key. The
min
setting (undertemperature
protection) is automatically activated.
ALARM
max
auto
40
0.0
K
min
15
.0
°C
.0°C
+-
1. By turning the turn control, adjust the
desired lower alarm limit value, in the
example on the right 35.5 °C.
The lower alarm limit value cannot be set
higher than the top one. If no undertem-
perature protection limit is required, set
the lowest temperature.
ALARM
0.0
K
min
°C
auto
max
°C
35
5
.
40
0
.
+-
2. Press the confirmation key to confirm.
The
max
display (overtemperature pro-
tection) is activated.
ALARM
0.0
K
min
°C
auto
max
°C
35
5
.
40
0
.
+-
3.
By turning the turn control, adjust the
desired upper alarm limit value, in the
example on the right 38.5 °C.
The monitoring temperature must be set
sufficiently high above the maximum set
temperature. We recommend 1 to 3 K.
ALARM
0.0
K
min
°C
auto
max
°C
35
.5
38
5
.
+-
4. Accept the upper alarm limit value by
pressing the confirmation key. The setting
of the automatic temperature monitor
(ASF) is automatically activated (
auto
).
ALARM
max
38
.5°C
0.0
K
min
35
.5°C
auto
+-
5. With the turn control, select ON (
) or
OFF (
).
ALARM
max
38
.5°C
0.0
K
min
35
.5°C
auto
+-