46
D33464 | Date 09/2018
Menu mode
7.3.6
Balance
The distribution of the heating/cooling power (balance) between the upper and lower Peltier
elements can be corrected depending on the specific application. The adjustment range is
from -50 % to +50 %.
upper heating/cooling power
upper heating/cooling power
lower heating/cooling power
-20%
lower heating/cooling power
+30%
SETUP
Balance
-20
%
2/2
SETUP
Balance
+30
%
2/2
Fig. 29
Distribution of the heating/cooling power (example): The –20 % (left) setting causes
the lower Peltier elements to work at 20 % less power than the upper ones. The +30 % (right)
setting causes the lower Peltier elements to work at 30 % more power than the upper ones.
The 0 % setting restores the default distribution settings.
7.3.7
Remote control
In the setup entry
Remote control
, you can set whether the
appliance should be controlled via remote control and if
so, in which mode. These settings are available:
►
Off
►
Read Only
►
Write
►
Write+Alarm
When the appliance is in remote control mode, the
symbol appears in the temperature display. In the
settings
Write
and
Write+Alarm
, the appliance cannot
be controlled at the ControlCOCKPIT until the remote
control has been switched off (setting Off) or set to
Read
Only
.
In order to use the remote control function, program-
ming skills and special libraries are required.
Setup
Balance
Remote Control
Off
Gateway 192.168.5 .1
+30 %
2/2
23
.2°C
TEMP
Set 38
.0°C