background image

D33464 | Date 09/2018  

15

Construction and description

2.10  Ambient conditions

 

The appliance may only be used in enclosed areas and under the following ambient condi-

tions:

 Ambient temperature

16 

º

C to 40 

º

C

Humidity rh

max. 70 % non-condensing

Overvoltage category

II

Pollution degree

2

Altitude of installation

max. 2,000 m above sea level

 

The appliance may not be used in areas where there is a risk of explosion. The ambient air 

must not contain any explosive dusts, gases, vapours or gas-air mixtures. The appliance is 

not explosion-proof.

 

Heavy dust production or aggressive vapours in the vicinity of the appliance could lead to 

sedimentation in the interior and, as a consequence, could result in short circuits or dam-

age to electrical parts. For this reason, sufficient measures to prevent large clouds of dust 

or aggressive vapours from developing should be taken.

2.11  Scope of delivery

 

Power cable

 

Tilt protection (see page 19)

 

Sliding grids (load capacity 30 kg each)

 

USB storage medium with software and  AtmoCONTROL manual

 

the operating instructions at hand

 

Calibration certificate

 

Water tank with connection hose

2.12  Optional  accessories

 

USB to Ethernet converter (Fig. 7). Makes it 

possible to connect the Ethernet connection 

interface (see page 11) to the USB port of 

a computer / laptop. 

Fig. 7 

 USB to Ethernet converter

Содержание hpp 1400

Страница 1: ...www memmert com www atmosafe net 100 ATMOSAFE MADE IN GERMANY HPP OPERATING INSTRUCTIONS CONSTANT CLIMATE CHAMBER HPP 1400 2200 ...

Страница 2: ...979 E mail service memmert com When contacting customer service always quote the product serial number on the nameplate see page 12 Shipping address for repairs Memmert GmbH Co KG Kundenservice Willi Memmert Str 90 96 DE 91186 Büchenbach Germany Please contact our customer service before sending appliances for repair or before returning equipment otherwise we have to refuse acceptance of the shipm...

Страница 3: ...ividual configurations and sizes illustrations in this manual may be slightly different from the actual appearance Function and operation are identical Other documents that have to be observed For operation of the appliance with MEMMERT AtmoCONTROL observe the separate software manual To open the AtmoCONTROL software manual click on Help in the AtmoCONTROL menu bar for service and repair work see ...

Страница 4: ...2 6 Designation nameplate 12 2 7 Technical data 13 2 8 Applied directives and standards 14 2 9 Declaration of conformity 14 2 10 Ambient conditions 15 2 11 Scope of delivery 15 2 12 Optional accessories 15 3 Delivery transport and setting up 16 3 1 Safety regulations 16 3 2 Delivery 16 3 3 Transport 16 3 4 Unpacking 16 3 5 Storage after delivery 17 3 6 Setting up 17 4 Putting into operation 21 4 1...

Страница 5: ...r failure 40 7 Menu mode 41 7 1 Overview 41 7 2 Basic operation in menu mode using the example of language selection 42 7 3 Setup 43 7 4 Date and Time 47 7 5 Calibration 48 7 6 Program 53 7 7 Sound 54 7 8 Protocol 55 7 9 User ID 56 8 Maintenance and service 57 8 1 Cleaning 57 8 2 Regular maintenance 58 8 3 Repairs and service 58 9 Storage and disposal 59 9 1 Storage 59 9 2 Disposal 59 Index 60 ...

Страница 6: ...er you or somebody else could be injured if you do not observe the accompanying safety regulation Caution is used for information that is important for avoiding damage 1 1 2 Signs used Warning signs warning of a danger Danger of electrocution Danger of explosion Dangerous gases vapours Danger of top pling over Hazard area Observe the opera ting instructions Prohibition signs forbidding an action D...

Страница 7: ...ning If the door is open while the appliance is in operation the appliance may overheat and pose a fire hazard Do not leave the door open during operation Warning In case of appliances of a certain size you can get accidentally locked in which is life threatening Do not climb into the appliance 1 3 Requirements of the operating personnel The appliance may only be operated and maintained by persons...

Страница 8: ...teri als the solvents of which could form an explosive mixture when combined with air If there is any doubt as to the composition of materials they must not be loaded into the appliance Potentially explosive gas air mixtures must not form neither in the working chamber nor in the direct vicinity of the appliance 1 6 Changes and alterations No unauthorised changes or alterations may be made to the ...

Страница 9: ... 2 Construction 1 ControlCOCKPIT with capacitive function keys and LCD displays see page 25 2 On Off switch see page 22 3 Door handle see page 23 4 Slide in units 5 Heated full sight glass door 6 Lockable castors with extendable feet see page18 7 Nameplate see page 12 8 Door handle see page 23 9 USB interface see page 11 ...

Страница 10: ...ad for the appliance must be carefully checked for chemical compatibility with the materials mentioned A material resistance table can be requested from the manufacturer 2 4 Electrical equipment Operating voltage and current consumption See nameplate Protection class I i e operating insulation with PE conductor in accordance with EN 61010 Protection type IP 20 acc to EN 60 529 Interference suppres...

Страница 11: ... located on the right side of the ControlCOCKPIT Fig 3 Ethernet interface Via Ethernet interface the appliance can be connected to a network so that you can trans fer programmes created with the AtmoCONTROL software to the appliance and read out protocols The Ethernet interface is located on the rear of the appliance Fig 4 For identification purposes each appliance con nected must have its own uni...

Страница 12: ...ont of the appliance on the right behind the right door see page 9 Typ HPP 1400 F Nr 0109 0088 230 V 13 5 A 50 60 Hz 3100 W DIN12880 Kl 3 1 Nenntemp 60 C 1 2 3 4 5 6 10 8 9 7 Fig 5 Nameplate example 1 Type designation 2 Operating voltage 3 Applicable standard 4 Protection type 5 CE conformity 6 Address of manufacturer 7 Disposal note 8 Temperature range 9 Connection power ratings 10 Appliance numb...

Страница 13: ...60 2140 Net weight kg 450 493 Current consumption A 230 V 50 60 Hz 13 5 15 2 115 V 50 60 Hz 27 30 4 Power W 3100 3500 max number of sliding grids 28 2 x 14 42 3 x 14 max load per sliding grid kg 30 max load per appliance kg 250 Temperature Adjustment range 15 C to 60 C2 Adjustment precision 0 1 K Humidity Adjustment range rh 15 to 80 1 see Fig 6 2 The minimum temperature depends on the outdoor tem...

Страница 14: ...1326 A2 2003 Directive 2006 95 EC amended Directive of the council for harmonisation of the laws of member states relating to electrical equipment designed for use within certain voltage limits Standards complied with DIN EN 61 010 1 VDE 0411 part 1 2002 08 DIN EN 61 010 2 010 VDE 0411 part 2 010 2004 06 EN 61 010 1 2001 EN 61 010 2 010 2003 2 9 Declaration of conformity You can download the EC de...

Страница 15: ... or aggressive vapours in the vicinity of the appliance could lead to sedimentation in the interior and as a consequence could result in short circuits or dam age to electrical parts For this reason sufficient measures to prevent large clouds of dust or aggressive vapours from developing should be taken 2 11 Scope of delivery Power cable Tilt protection see page 19 Sliding grids load capacity 30 k...

Страница 16: ... in a wooden crate and delivered on a wooden pallet 3 3 Transport The appliance can be transported in three ways With a forklift truck move the forks of the truck entirely under the pallet On a manual pallet jack on its own castors for which the catch on the front castors must be released 3 4 Unpacking To avoid damage do not unpack the appliance until you reach the installation site Unscrew wooden...

Страница 17: ...t the Memmert service see page 2 3 6 1 Preconditions The appliance may only be installed on the bottom The installation site must be flat and horizontal and must be able to reliably bear the weight of the appliance see Technical data on page 13 Do not place the appliance on a flammable surface A suitable power source see type plate and technical data on page 13 must be available at the installatio...

Страница 18: ... the mechanism pull out the ratchet lever Fig 10 To retract and extend the stand actuate the ratchet lever The direction of movement up down can be adjusted using the rocker above the ratchet lever Press in the rocker on the right side to extend the stand with the ratchet lever Fig 11 Press in the rocker on the left side to retract the stand with the ratchet lever Fig 12 Fig 11 To extend the stand...

Страница 19: ...on The tilt protection is included in the delivery 3 Screw the tilt protection onto the back of the ap pliance as illustrated 4 Bend the tilt protec tion upwards by 90 in the desired distance to the wall observe the minimum distance to the wall see Fig 8 5 Drill a hole insert a dowel and screw the tilt protection to a suitable wall ...

Страница 20: ...o adjusting screws each at the top and the bottom of each door for this purpose Fig 13 First adjust the setting at the top of the door and if this is not sufficient adjust the screws at the bottom of the door 1 Open the door 2 Loosen the screws 3 Adjust the position of the door 4 Tighten the screws again 5 Check the position of the door 6 Readjust if required Fig 13 Adjusting the doors ...

Страница 21: ...ble so that it is easily accessible at all times and can be pulled off quickly for example in case of interference or an emergency it does not represent a trip hazard it cannot come into contact with any hot parts 4 2 Filling and connecting the water tank Water specifications For steam creation use only demineralised desalinated water Aqua Dem with a conduc tivity 10 μS cm Otherwise possible calci...

Страница 22: ...by three animated white dots If the dots have another colour an error has occurred see page 39 After the first start up the appliance display is set to English by default You can change the language as described from page 42 However to get a basic overview of operating the appliance you should read the following chapter first ON O O ON N N N O ON O Fig 16 Switching on the appliance ...

Страница 23: ...d closing the doors To open the doors pull the door handle away from the appliance to the left or to the right depending on the door variation see Fig 17 and open it To close the appliance push the door closed and push the door handle to the front Fig 18 Fig 17 Opening the door Fig 18 Closing the door Warning Leaving the door open during operation can cause the appliance to overheat and pose a fir...

Страница 24: ... materials of the appliance see page 10 Insert the sliding steel grids or sliding shelves The maximum number or grids shelves and the load capacity are specified in the technical data overview from page 13 The chamber must not be loaded too tightly so that proper air circulation in the interior is guaranteed Do not place any of the chamber load on the bottom touching the side walls or right below ...

Страница 25: ... 19 14 15 16 17 9 10 11 12 13 TIMER 30m 04h End 13 30 23 11 Manual mode 12 09 2012 13 44 7 Fig 20 ControlCOCKPIT of HPP 1400 2200 appliances in operating mode 1 Activation key for temperature setpoint adjustment 2 Setpoint and actual temperature display 3 Fan speed display 4 Activation key for fan speed setting 5 Switch to menu mode see page 41 6 Activation key for the appliance state 7 Appliance ...

Страница 26: ...he former values If you want to cancel the setting procedure press the activation key on the left or right of the display that you want to exit The appliance restores the former values Only the settings that you have confirmed by pressing the confirmation key before cancelling the setting procedure are accepted 5 4 3 Operating modes The appliance can be operated in different modes Manual mode The ...

Страница 27: ...djustment options As described in chapter 5 4 2 you can set the following parameters after pressing the cor responding activation key in any sequence Temperature Adjustment range 15 C to 60 C Heating operation is indicated by the symbol Cooling is indicated by the symbol You can select C or F as the temperature unit displayed see page 44 22 4 37 0 TEMP Set C C The minimum temperature that can be r...

Страница 28: ...m hours minutes format For 24 hours and more the format dd hh days hours is used The maximum duration adjustable is 99 days 00 hours 3 Press the confirmation key to confirm The display now shows the remaining time in a large font and the approximate end time in a smaller font beneath The status display shows Timer active TIMER 30m 04h End 13 30 23 11 Timer active 12 Sept 2012 13 44 4 Now as descri...

Страница 29: ...display The current operating mode is highlighted automatically in this example Manual mode Holz trocknen aufheizen 09 12h Fr 20 10 2010 20 31 min ALARM of C max 000 C rh C GRAPH off ON LIGHT ON LIGHT Manueller Modus 12 09 2012 13 44 ALARM max 38 5 C min 36 5 C Manual mode 12 09 2012 13 44 Activate 2 Turn the turn control until the start symbol is highlighted The current programme is displayed in ...

Страница 30: ...ess the confirmation key to confirm The programme is cancelled 13 44 End Test 012 12 11 2012 A cancelled programme cannot be resumed at the point it was cancelled It must be restarted from the beginning End of programme End is shown on the display to indicate that the programme has finished 13 44 End Test 012 12 11 2012 You can now restart the programme as described select another programme for pr...

Страница 31: ... activated in menu mode Sound see page 54 indicated by the speaker symbol in the alarm display the alarm is additionally signalled by an intermittent acoustic signal which can be turned off by pressing the confirmation key Information on what to do in this case is provided in chapter Malfunctions warning and error messages from page 37 Before reading how to adjust temperature monitoring from page ...

Страница 32: ...erature violates the set tolerance band around the setpoint in the example in Fig 23 50 C 2 K e g if the doors are opened during operation section B of illustration the alarm is set off The ASF alarm is automatically terminated as soon as 50 of the set toler ance band of the setpoint in the example 50 C 1 K are reached again section C If the temperature setpoint is altered the ASF is automatically...

Страница 33: ...han the top one If no undertem perature protection limit is required set the lowest temperature ALARM 0 0K min C auto max C 355 40 0 2 Press the confirmation key to confirm The max display overtemperature pro tection is activated ALARM 0 0K min C auto max C 355 40 0 3 By turning the turn control adjust the desired upper alarm limit value in the example on the right 38 5 C The monitoring temperatur...

Страница 34: ...ed in menu mode Sound see page 54 as indicated by the speaker symbol the alarm is additionally signalled by an intermittent acoustic signal Information on what to do in this case is provided in chapter Malfunc tions warning and error messages from page 37 Adjusting humidity monitoring 1 Press the activation key to the left of the ALARM display The temperature moni toring setting is automatically a...

Страница 35: ...ing the turn control adjust the desired upper alarm limit in the example on the right 70 rh ALARM min 50 0 rh max 70 0 rh 7 Accept the selection by pressing the confirmation key and leave the Alarm display by pressing the activation key on the side Humidity monitoring is now active ALARM min 50 0 rh max 70 0 rh 5 6 Graph The GRAPH display provides an overview of the chronological sequence of the s...

Страница 36: ...e 1 Activate graphic representa tion as described above and then press the activation key next to the parameter selec tion 0 4 8 12 16 20 24 C 40 20 60 80 100 Fr 20 10 2010 20 12 09 2012 14 00 16 00 18 00 38 39 40 2 Select humidity with the turn control C 40 20 60 80 100 Fr 20 10 2010 20 12 09 2012 38 39 40 3 Press the confirmation key to confirm The humidity profile is displayed You can change th...

Страница 37: ...indicated by the speaker symbol the alarm is additionally signalled by an intermittent acoustic signal If the confirmation key is pressed the acoustic alarm can be temporarily switched off until the next alarm event occurs 6 1 1 Temperature monitoring Description Cause Action See Temperature alarm and ASF are displayed TEMP ASF Set 38 5 C 40 4 C Automatic tem perature monitor ASF triggered Check i...

Страница 38: ...Check the water supply and the filling level of the water tank If required refill the water tank If the error occurs again contact customer service page 21 page 2 6 2 Malfunctions operating problems and appliance errors Error description Cause of error Rectifying errors See Displays are dark External power supply was interrupted Check the power supply page 21 Miniature fuse appliance fuse or power...

Страница 39: ...emporarily be kept in service Contact customer service as soon as possible page 2 Error message E 3 in the temperature display TEMP Set 45 0 C E 3 C Operating and monitoring sensor defective Switch off appli ance Remove the cham ber load Contact customer service page 2 Error message E 6 in the humidity display HUMIDITY Set 50 0 rh rh E 6 Humidity sensor defective No humidity control possible Conta...

Страница 40: ...ion of the power failure are documented in the log memory In timer or programme mode In case of an interruption of the power supply of less than 60 minutes the current pro gramme is continued from the point at which it was interrupted For longer interruptions of the power supply all appliance functions heating fan etc are switched off In remote control mode The previous values are restored If a pr...

Страница 41: ...SIGNALTÖNE PROG ZEIT UND DATUM CALIB USER ID PROTOCOL LANGUAGE TIME SOUND ON O O ON N N N O ON O 1 2 3 4 5 12 14 15 13 6 7 18 19 8 9 10 11 16 17 Fig 26 ControlCOCKPIT in menu mode 1 Language selection activation key 2 Language selection display 3 Date and time display 4 Date and time setting activation key 5 Exit menu mode and return to operating mode 6 Setup activation key basic appliance set tin...

Страница 42: ...tivation key which you have used to activate the display The appliance returns to the menu overview Only the settings that you have con firmed by pressing the confirmation key before cancelling the setting procedure are accepted SETUP SIGNALTÖNE CALIB LANGUAGE TIME ON O O ON N N N O ON O 2 With the turn control select the desired new setting e g Español Spanish 3 Save the setting by pressing the c...

Страница 43: ...6 Remote control see page 46 Gateway see page 47 If the SETUP menu contains more entries than can be displayed this is indicated by the display 1 2 This means that there is a second page of entries To display the hidden entries use the turn control to scroll beyond the lowest entry The page display changes to 2 2 7 3 2 IP address and subnet mask If you want to operate one ore more appliances in a ...

Страница 44: ...t IP address 255 168 100 100 Subnet mask 255 255 0 0 Alarm temp C F TWW TWB Timer mode 4 Accept the selection by pressing the confirmation key The next three digits of the IP address are automatically selected Setting these is done according to the description above SETUP Unit IP address 255 168 100 100 Subnet mask 255 255 0 0 Alarm temp C F TWW TWB Timer mode 5 After setting the last three digits...

Страница 45: ...s right after activation B Timer setpoint dependent Timer does not start until tolerance band is reached 7 3 5 Type of the slide in unit Grid or Shelf Here you have to set the type of the slide in unit grid or shelf used The selection Shelf enables you to adjust the control function to the different air flow characteristics in the interior when using optional sliding shelves instead of the grids t...

Страница 46: ...e lower Peltier elements to work at 30 more power than the upper ones The 0 setting restores the default distribution settings 7 3 7 Remote control In the setup entry Remote control you can set whether the appliance should be controlled via remote control and if so in which mode These settings are available Off Read Only Write Write Alarm When the appliance is in remote control mode the symbol app...

Страница 47: ...right side of the Time display The display is enlarged and the first adjustment option Date automatically highlighted Date and time Time Time zone Daylight savings 12 05 2012 12 00 GMT 01 00 Date 2 Turn the turn control until Time zone is highlighted Time Daylight savings 12 05 2012 12 00 GMT 01 00 Date Time zone 3 Accept the selection by pressing the confirmation key Time Daylight savings 12 05 2...

Страница 48: ...rs minutes in the same way Accept the selection by pressing the confirma tion key Time Time zone Daylight savings 2012 12 00 GMT 00 00 Date 27 05 7 5 Calibration To guarantee perfect control we recommend to calibrate the appliance once a year 7 5 1 Temperature adjustment The appliances are temperature calibrated and adjusted at the factory In case readjustment should be necessary later on for exam...

Страница 49: ...USTIEREN Temperatur letzte Justierung 12 10 2012 12 00 40 0 C K 0 2 C K 0 1 C K 0 2 Calibration Last updated 12 10 2012 12 00 00 Cal1 Cal2 Cal3 20 0 40 0 5 0 Temperature Humidity 2 Press the confirmation key repeatedly until the calibration temperature Cal2 is selected C K 0 2 C K 0 1 C K 0 2 e Cal1 Cal2 Cal3 40 0 5 0 20 0 3 With the turn control set the calibration temperature Cal2 to 30 C C K 0 ...

Страница 50: ...rument for example displays 31 6 C 30 0 C TEMP Set 30 0 C 31 6 C 9 In the SETUP adjust the calibration value Cal2 to 1 6 K actual value measured minus setpoint temperature and save the setting by pressing the confirmation key C K 0 2 C K C K 0 2 e Cal1 Cal2 Cal3 40 0 5 0 30 0 1 6 10 After the calibration procedure the temperature measured by the reference instrument should now also be 30 C 30 0 C ...

Страница 51: ...e activation key to the right of the CALIB display The display is enlarged and the temperature adjustment option is automatically highlighted Cal1 C 0 2K Cal2100 0 C 0 1K Cal3180 0 C 0 2K JUSTIEREN Temperatur letzte Justierung 12 10 2012 12 00 40 0 C K 0 2 C K 0 1 C K 0 2 Calibration Last updated 12 10 2012 12 00 00 Cal1 Cal2 Cal3 20 0 40 0 5 0 Temperature Humidity 2 Turn the turn control until Hu...

Страница 52: ... in the appli ance s working chamber 8 Close the door and in manual mode adjust the set humidity to 60 rh HUMIDITY 29 5 rh Set 60 0 rh 9 Wait until the appliance reaches the set humidity and displays 60 rh The reference instrument for example displays 58 5 rh HUMIDITY 60 0 rh Set 60 0 rh 58 5 rh 10 In the SETUP adjust the calibration value Cal2 to 1 5 actual value measured minus setpoint humidity ...

Страница 53: ...Select function by pressing the confirmation key All programmes available are displayed including the ones saved on the USB storage medium identified by the USB symbol The programme currently available for use is highlighted in orange Select Delete Test 012 Test 022 Test 013 Test 014 Test 015 Test 022 3 With the turn control select the pro gramme you want to make available for use Select Delete Te...

Страница 54: ...e shown on Sound Keysound At the end If door open On alarm If you want to edit another list entry Turn the turn control until the respective entry e g If door open Special accessories is highlighted in colour Keysound At the end If door open On alarm 2 Save the selection by pressing the confir mation key The adjustment options are automatically highlighted Keysound At the end If door open On alarm...

Страница 55: ...ce is not modified or deleted by reading it out 1 Connect the USB storage medium to the interface on the right side of the control panel 2 Activate the protocol To do so press the activation key on the right side of the PROTOCOL display The display is enlarged and the period This Month au tomatically highlighted To select another logging period use the turn control Protocol 12 10 2012 12 00 contro...

Страница 56: ... AtmoCONTROL manual 7 9 2 User ID activation and deactivation 1 Connect the USB storage medium with the user ID data to the interface on the right side of the control panel 2 Activate the user ID To do so press the activation key on the right side of the User ID display The display is enlarged and the entry Activate automatically highlighted USER ID Deactivate Activate 3 Confirm the activation by ...

Страница 57: ...ts come into contact with the working chamber or with the stainless steel housing Rust deposits can lead to an infection of the stainless steel If rust spots should appear on the surface of the working chamber due to impurities the affected area must be immediately cleaned and polished 8 1 2 Plastic parts Do not clean the ControlCOCKPIT and other plastic parts of the appliance with caustic or solv...

Страница 58: ...e perfect control we recommend to calibrate the appliance once a year see page 48 8 3 Repairs and service Warning After removing covers live parts may be exposed You may receive an electric shock if you touch these parts Disconnect the mains plug before removing any covers Any work inside the appliance may only be performed by qualified electri cians Repairs and service work are described in a sep...

Страница 59: ...e into their national laws It may not be disposed of in normal household waste For disposal please contact your dealer or the manufacturer Any appliances that are infected infectious or contami nated with materials hazardous to health are excluded from return Please also observe all other regulations applicable in this context Before disposing of the appliance please render the door locking mechan...

Страница 60: ...n speed 27 Forklift truck 16 Fresh water container 21 G Gateway 47 Graph 35 Grid 45 H Hazards 7 Heat sink 57 Height adjustment 18 Humidity 27 Humidity adjustment 51 Humidity calibration 51 Humidity monitoring 34 35 38 Humidity profile 36 I Installation site 17 Intended use 8 Interfaces 10 IP address 43 L Language selection 42 Loading the appliance 24 Log memory 40 55 M Maintenance 57 Malfunctions ...

Страница 61: ...on 49 Temperature monitor 31 32 Temperature monitoring 30 31 Temperature profile 35 Temperature sensor 31 Tilt protection 19 Time 45 Timer 28 Timer mode 45 Transport 16 Transport damage 16 Turn control 26 TWW 31 TWW temperature monitor ing 32 U Unit 44 Unpacking 16 USB interface 11 55 User ID 56 W Warning messages 11 37 Water specifications 21 Water tank 21 Weight 13 ...

Страница 62: ......

Страница 63: ......

Страница 64: ...rt Straße 90 96 D 91186 Büchenbach Tel 49 9122 925 0 Fax 49 9122 14585 E Mail sales memmert com facebook com memmert family Die Experten Plattform www atmosafe net Constant Climate Chamber HPP 1400 2200 D33464 Date 09 2018 englisch ...

Отзывы:

Похожие инструкции для hpp 1400