D33464 | Date 09/2018
27
Operation and control
The status display shows you which operating mode or operating state the appliance is
currently in. The current operating state is highlighted in colour and indicated by the text
display:
Appliance is in programme mode
■
Programme is stopped
Appliance is in manual mode
The example on the right shows the appliance
in manual mode, identified by the coloured
hand symbol.
►
When the appliance is in timer mode,
Timer active
is displayed:
►
When the appliance is in remote control mode, the
symbol appears in the temperature display:
5.4.4
Manual mode
In this operating mode, the appliance runs in permanent operation at the values set on the
ControlCOCKPIT.
Adjustment options
As described in chapter 5.4.2 , you can set the following parameters after pressing the cor-
responding activation key (in any sequence):
Temperature
Adjustment range: 15 °C to 60 °C
Heating operation is indicated by the symbol.
Cooling is indicated by the symbol.
You can select °C or °F as the temperature unit displayed (see
page 44).
22
.4
37
.0
TEMP
Set
°C
°C
The minimum temperature that can be reached depends on the surrounding conditions.
The devices can cool down to 10 K below room temperature. For this purpose, the Peltier
module needs sufficient ventilation (see Fig. 8 on page 17).
Fan speed
Adjustment range: 10 to 100 % in steps of 10%
FAN
%
50
FA
FA
N
N
Humidity
Adjustment range: 15 to 80 % rh
Humidification is indicated by the symbol.
Dehumidification is indicated by the symbol.
HUMIDITY
70
%rh
Set
34
% rh
Manual mode
12.09.2012
13:44
Timer active
12.Sept.2012 13:44
23
.2°C
TEMP
Set 38
.0°C