D33464 | Date 09/2018
45
Menu mode
7.3.4
Timer mode
Here, you can choose whether the digital backwards counter with
target time setting (see page 28) should run setpoint-de-
pendent or not. This determines whether the timer should not
start until a tolerance band of ±3 K around the set temperature
is reached (Fig. 28, B) or if it should start right after activation
(A).
t
Timer
t
Timer
A
B
°C/°F
t
6 K
Fig. 28
Timer Mode
A
Timer independent of setpoint: Timer starts right after activation
B
Timer setpoint-dependent: Timer does not start until tolerance band is reached
7.3.5
Type of the slide-in unit (
Grid
or
Shelf
)
Here, you have to set the type of the slide-in unit (grid or shelf)
used. The selection Shelf enables you to adjust the control
function to the different air flow characteristics in the interior
when using optional sliding shelves instead of the grids that are
part of the standard delivery.
Unit
IP address
255.145.136.225
Subnet mask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer mode
S i
i
i
Unit
IP adress
255.145.136.225
Subnet mask
255.255.0.0
°C
°F
Timer mode
Alarm temp
TWW
TWB
Slide-in unit
Grid Shelf