34
D33464 | Date 09/2018
Operation and control
6. Press the confirmation key to confirm.
The ASF tolerance band setting is acti-
vated.
ALARM
K
min
°C
auto
max
38
.
5°C
35
5
.
0.0
+-
7.
With the turn control, adjust the desired
tolerance band, e.g. 2.0 K.
We recommend a tolerance band of 1 to
3 K.
ALARM
K
min
°C
auto
max
38
.
5°C
35
5
.
1.0
+-
8. Press the confirmation key to confirm.
Temperature monitoring is now active.
ALARM
max
38
.5°C
1.0
K
min
35
.5°C
auto
+-
5.5.2
Humidity monitoring
If humidity monitoring was triggered, this
is indicated by the humidity display: the
actual humidity is highlighted in red and a
warning symbol is shown ( Fig. 24 ). If the
acoustic alarm has been activated in menu
mode (
Sound
, see page 54, as indicated
by the speaker symbol
), the alarm is
additionally signalled by an intermittent
acoustic signal. Information on what to do
in this case is provided in chapter Malfunc-
tions, warning and error messages from
page 37.
Adjusting humidity monitoring
1. Press the activation key to the left of the
ALARM
display. The temperature moni-
toring setting is automatically activated.
min
000°C
ALARM
max
000
°C
auto off
99K
-
+
ALARM
max
40
.0°C
auto
0.0
K
min
15
.0°C
+-
2. Turn the turn control until the humidity
monitoring entry is highlighted.
ALARM
min
40
.0%rh
max
60
.0%rh
min
000°C
ALA
ON
LIGHT
ON
ON
ON
LIGH
IGH
IGH
T
T
T
ON
LIGHT
FEUCHTE
30
.0%rh
Set 30
.0%rh
°C
2
.4
°C
t 37.0
LIC
1
AL
auto
min
35
.5
HUMIDITY
Set 70
.0%rh
%rh
75
.4
m
FEUCHTE
30
.0%rh
Set 30
.0%rh
C
°C
Set 37.0
m
0
HUMIDITY
Set 70
.0%rh
%rh
75
.4
Fig. 24
Humidity monitoring triggered