The
Trigger Source Menu
allows the user the flexibility to set up the Start and Stop
trigger to the desired configuration, that being NO contacts, NC contacts, Voltage
Applied or Voltage Removed. After making the desired selections press the OK
function key to initialize the new configuration and return to the High Speed Capture
screen.
The Power MultiMeter is an auto ranging unit but if the user prefers to set up the
Current and Voltage to the expected measurement range just simply press the
Change
I & V Range
function key and the following screen will appear.
IMPORTANT
Default ranges are 750 Volts and 100 Amps. This mode allows the
user to select and set the voltage and current ranges before initalizing the measurement
test. All ranges have a +20% of the selected range, if the amplitude exceeds the +20%,
49