THREE PHASE WYE:
Three Phase Wye screen:
m3y;
query the readings:
qr;
Three Phase mode phase angle reference source selection read:
s3pphrefr;
sends ( phref = 0 ) in 0-5 integer format. 0=Va, 1=Vb, 2=Vc, 3=Ia, 4=Ib, 5=Ic
Three Phase mode phase angle reference source selection write:
s3pphrefw,1;
Note
: once phase angle reference source setting is changed the change is not
recognized until 3 phase mode is re-entered (i.e. ‘m3y’ or ‘m3d’ command is re-issued).
Changes rms/average voltage setting:
srmsavgw,K;
K= 0 for rms, K=1 for avg.
Lock keypad:
keylock;
Three Phase Wye screen and with automatic interval reading query:
m3yi,K;
( K = interval in minutes, 1 minute minimum to 60 minutes maximum ).
(
Note
: when any other command is executed in this mode the interval query operation
is canceled.)
Unlock keypad:
keyunlock;
THREE PHASE DELTA:
Three Phase Delta screen:
m3d;
query the readings:
qr;
Three Phase mode phase angle reference source selection write:
s3pphrefw,1;
Note
: once phase angle reference source setting is changed the change is not
recognized until 3 phase mode is re-entered (i.e. ‘m3y’ or ‘m3d’ command is re-issued).
Changes rms/average voltage setting:
srmsavgw,K;
K= 0 for rms, K=1 for avg.
Lock keypad:
keylock;
Three Phase Delta screen and with automatic interval reading query:
m3di,K;
( K = interval in minutes, 1 minute minimum to 60 minutes maximum ).
(
Note
: when any other command is executed in this mode the interval query operation
is canceled.)