Select
CHANGE I &
V
RANGE
.
-
Press
CHANGE I RNG
to set the current range desired.
-
Press
CHANGE V RNG
to set the voltage range desired.
6.
To enter the new configuration and return to the previous screen,
press the OK function key.
7.
Once
all
the connections to the instrument are made, press the
ARM
function key to arm the capture circuitry.
NOTE:
Status will change from
INACTIVE
to
ARM
.
8.
Use
the
START
inputs to initiate the test. If a BEFORE STOP
mode of operation is selected, a
STOP
signal is required to
terminate the test. In a AFTER-START mode, the test is
automatically terminated depending on selection: 15, 30 or 60
seconds after it is started (15 sec/300 samples, 30 sec/600
samples or 60 sec/1200 samples).
The
START
or
STOP
inputs respond to a change of state. This
change may be a dry contact opening or closure, initial application
of an AC or DC voltage and . The input terminals are isolated from
the chassis ground and are independent of signal polarity.
9.
Once the measurement is completed and data has been captured,
the data is automatically displayed. Use the
PAGE UP
or
PAGE
DOWN
function keys to scroll through the pages of data. The total
samples of readings and number of pages is dependent on
selection; 15 sec/300 samples, 30 sec/600 samples or 60 sec/1200
samples.
26