If a High Speed measurement is desired while in the Three Phase Wye or Three Phase
Delta just simply press the
High Speed Capture
function key and the following screen
will be displayed.
The
High Speed Capture
screen has numerous function key options that allows the
user to make before conducting the high speed measurement.
•
Trigger Input Select
-make changes to the Timer start and stop binding posts; NO
or NC contacts, Voltage Applied and Voltage Removed.
•
Change I & V Range
- change the current and voltage range desired for each
particular measurement.
•
Capture Mode
-After Start or Before Stop.
•
Arm
-arms the Power MultiMeter for immediate measurement.
•
Select Sample Number
-allows selection of; 15 sec/300 samples, 30 sec/600
samples, 60 sec/1200 samples.
•
ESC
- selection of this function key recalls the previous screen.