21
- From
the
Main Menu, press the More function key.
-
From the More screen, press the SetUp Page2 function key.
-
From the SetUp Menu screen, press the
⇑
or
⇓
Arrow keys
to cursor to the preset CT ratios or the Customized Option.
-
Preset ratioes are:
None
or
1:1, 1000:5 (200-1), 3000-5
(600-1), 1000:1, 3000:1,
Customized, CT and PT.
-
In
the
Customized SetUp screen, the user can enter any
ratio up to 9999:1 or 9999:5.
-
Once the selection is entered, press the OK function key to
return to the previous screen.
-
Make the proper CT clamp connections to the unit and select
the type of measurement (Single or Three Phase) desired.
-
The Power MultiMeter will measure the current or currents
by the selected ratio or
CUSTOMIZED
CT or PT ratio.
-
After completion of the CT clamp on measurements, it is
recommended that the unit be returned to its default setting:
(None or 1:1).
-
Return to the CT ratio screen by following the same steps as
previously described or simply turn
OFF
the Power
Multimeter.
3.
All measuring functions are explained in sections 4.5. and 4.6.
4.7.2 VERIFYING THE RATIOS OF CTS
1.
For this application, connect the clamp-on CT directly to the phase
A current channel terminals on the front panel. Connect the
secondary of the CT under test to the phase B current channel
terminals.