background image

AC Voltage (V

~

)

#61-340
#61-342

Digital Multimeter
Instruction Manual

Read First:  Safety Information

Understand and follow operating instructions
carefully. If this tester is not used in a manner
specified by IDEAL, protection provided by the
product may be impaired.

WARNINGS

To avoid possible electric shock, personal injury or
death follow these instructions: 
• Do not use if meter appears damaged.  
• Visually inspect the meter to ensure case is not

cracked and back case is securely in place.

• Inspect and replace leads if insulation is dam-

aged, metal is exposed, or probes are cracked.
Pay particular attention to the insulation sur-
rounding the connector.

• Do not use meter if it operates abnormally as

protection maybe impaired.  

• Do not use during electrical storms or in wet

weather.

• Do not use around explosive gas, dust, or vapor.
• Do not apply more than the rated voltage to the

meter. 

• Do not use without the battery and the back case

properly installed.

• Remove the test leads from the meter before

removing battery cap.

• Do not attempt to repair this unit as it has no

user-serviceable parts.

•  Disconnect power and discharge capacitors

before testing resistance, continuity, diodes,
capacitance or temperature.

• Replace battery as soon as low battery indicator    

appears to avoid false readings.

• Use the proper terminals, function and range for

your measurements.

• Comply with local and national safety

requirements, including the use of appropriate
personal protective equipment.

CAUTION:

To protect yourself, think "Safety First":
• Voltages exceeding 30VAC or 60VDC pose a

shock hazard so use caution.

• Use appropriate personal protective equipment

such as safety glasses, face shields, insulating
gloves, insulating boots, and/or insulating mats.

• Before each use:

- Perform a continuity test by touching the

test leads together to verify the functionality
of the battery and test leads.

- Use the 3 Point Safety Method.  (1) Verify

meter operation by measuring a known
voltage.  (2) Apply meter to circuit under
test.  (3) Return to the known live voltage
again to ensure proper operation.

• Never ground yourself when taking electrical

measurements.

• Connect the black common lead to ground or

neutral before applying the red test lead to
potential voltage.  Disconnect the red test lead
from the voltage first.

• Always work with a partner.
• When using the probes, keep fingers as far

behind the probe tips as possible.

Ranges & Accuracies:  

AC Converter:  

61-340 model is averaging

sensing, rms calibrated; 61-342 model is true rms
sensing. 

Accuracy:  

Accuracy is specified as +/-(a

percentage of the r a fixed amount) at
23°C±5°C (73.4°F ± 9°F), less than 75% relative
humidity.

Temperature Coefficient:

0.1 times the

applicable accuracy specification from 32°F to
64°F and  82°F to 122°F (0°C to 18°C ; 28°C to
50°C).

Ranges & Accuracies

DC Voltage (V )

DC

+

Toggle for
Frequency
(>20% of
range)

Function

Range &

Resolution

Accuracy

Overload Protection

DC

V

oltage

400.0m/4.000/40.00/400.0 V

±(0.5%+5)

600.0V

±(1.0%+3)

AC V

oltage

400.0m/4.000/40.00/400.0/600 V

±(1.5%+5)

(40~400Hz)

DC

Current

400.0/4000 µA ; 40.00/400.0 mA

±(0.5%+5)

500mA/600V  Fast Fuse

4.000/10.00 A

±(1.5%+5)

10A/1000V  Fast Fuse

AC

Current

400.0/4000 µA

±(1.5%+5) 

500mA/600V  Fast Fuse

(40~400Hz)

40.00/400.0 mA

±(2.0%+5) 

4.000/10.00 A

±(2.5%+5) 

10A/1000V Fast Fuse

400.0

±(1.2%+5) 

Resistance

4.000k/40.00k/400.0k 

±(1.0%+2)

600V DC/AC rms

4.000M

±(1.2%+2)

40.00M

±(2.0%+5)

40.00nF

±(3.0%+10)

Capacitance*

400.0n/4.000µ/40.00µ

F

±(3.0%+5) 

600V DC/AC rms

400.0µ/4000µ

F

±(20%+5) 

Frequency

10.00/100.0/1.000k/10.00k/100.0k/1.000M/10.00M Hz

±(0.1%+3) 

600V DC/AC rms

Sensitivity: <1

MHz: 0.7V

rms ; >1MHz: 5V

rms 

Duty

Cycle

0.1 - 99.9%

±(2.5%+5)   (<10kHz)

Diode Check

Test current : (1±0.6) mA and then open circuit voltage is 2.5VDC typical.

600V DC/AC rms

Continuity

The beeper turns on <25

and turns off at >120

.

600V DC/AC rms

Temperature**

-58~1500°F

±(3.0%+5)

-50~800°C

±(3.0%+3)

*Accuracy not available for <10nF capacitance.   **Accuracy is stated for meter only

.  Thermocouple accuracy adds another ±2.5%

 to reading.

Input Impedance :  10M

for V

AC &

VDC.              CF > 2, add ±1.0% to accuracy

.

900VDC

750V

AC rms

+

~

Symbols

Risk of electric shock

See instruction card

DC measurement

Equipment protected by double or 
reinforced insulation

Battery

Earth

AC measurement

Conforms to EU directives

ND 6447-1 61-340 61-342 Ins  3/17/08  6:42 AM  Page 1

Отзывы: