43
9. Example:
Correction factor = [ ( Standard reading / Measured ( PMM1 ) reading - 1 ] x
1,000,000
3.00000 / 2.99811 - 1 x 1,000,000 = 630.397, the new correction factor to be
entered is a positive 630. Enter
cfw, 0, 630;
then update by typing
cfu;
. Inject
the same source and verify that the correction factor has brought the range in
question back into tolerance. Using the % error formula calculate the % error of
the readings.
Keypad
lock
keypad:
keylock;
unlock
keypad:
keyunlock;
keypad test mode:
keytest;
restore normal key operation after test mode:
keynormal;
Note:
A screen command (i.e. mpu, m1, m3y…) must be issued to restore
normal key operation.
General
V25 and DSP firmware revision level query:
grev;
returns revs in format: V25rev=K.k, DSPrev=L.1.
EXAMPLES
SINGLE PHASE:
Single Phase screen:
m1;
query the readings:
qr;
Single Phase custom input selection read;
s1pcustomr;
sends channel 1 and 2 input source selections in 0-8 integer format.
( chan1=0 chan2=6 )
0=Van, 1=Vbn, 2=Vcn, 3=Vab, 4=Vbc, 5=Vca, 6=Ia, 7=Ib, 8=Ic.
Single Phase custom input selection write:
s1pcustomw,m,n;
Write channel 1 and channel 2 input source selections ( same format as read)
m=channel 1, n=channel 2.
Note
: Once input source settings are changed the change
is not recognized until single phase mode is re-entered (i.e. ‘ml’ command is re-issued).
Changes rms/average voltage setting:
srmsavgw,K;
K= 0 for rms, K=1 for avg.
Lock keypad:
keylock;
Single Phase screen and with automatic interval reading query:
m1i,K;
( K = interval in minutes, 1 minute minimum, 60 minutes maximum ).
(
Note
: when any other command is executed in this mode the interval query operation
is canceled.)