Alarm
Problem
Solution
10 Patient Temperature
1 Low Medium Priority
Alarm
The Patient Temperature 1 reading
is below 31°C (87.8°F), and the water
temperature is below 31°C (87.8°F),
and the system is continuing to cool
the patient when the system is in a
patient control mode (e.g. Control
Patient, Cooling or Rewarming).
1. Verify Patient Temperature Accuracy
Confirm patient temperature using a secondary site.
Confirm that the primary temperature probe is properly placed
and registering an accurate temperature.
2. Verify Patient / Water Temperature Control
Observe patient temperature.
Is the patient temperature increasing?
Observe water temperature.
Is the water temperature >31°C (87.8°F)?
Is the water temperature increasing?
If the water and/or patient temperatures are beginning to
increase, the system is working correctly. Continue to monitor
patient temperature.
3. Verify Water Temperature Limits
In
Normothermia Settings
or
Hypothermia Settings
, confirm
that the water temperature high limit is set ≥36°C (96.8°F).
If necessary, set water temperature high limit ≥36°C (96.8°F).
Resume therapy.
Monitor water temperature and patient temperature.
If the water and/or patient temperatures are beginning to
increase, the system is working correctly. Continue to monitor
patient temperature.
4. Verify Water Control – System Performance
Place unit in Manual mode, with water temperature set
>38°C (100.4°F).
Monitor water temperature and patient temperature.
If the water temperature is not increasing to the commanded
temperature, the water heating system is not working. Call BD
Customer Support.
If the water temperature increases to the commanded
temperature, the water heating system is working.
5. Verify Patient Control – System Performance
Place system back into the patient control mode.
Monitor water temperature and patient temperature.
If the water and/or patient temperatures are continuing to
increase, the system is working correctly. Continue to monitor
patient temperature.
If the water temperature and/or patient temperatures are not
increasing, call BD Customer Support.
11 Patient Temperature
1 Below Low Patient
Low Priority Alarm
In Normothermia Therapy: The
Patient Temperature 1 reading is
below the Low Patient Alarm setting
in Normothermia Settings.
In Hypothermia Therapy: The Patient
Temperature 1 reading is below
the Low Patient Alarm setting in
Hypothermia Settings.
If desired, adjust the
Low Patient Alarm
setting in
Normothermia Settings
or
Hypothermia Settings.
12 Patient Temperature
1 High
Low Priority Alarm
The Patient Temperature 1 reading is
above 39.5°C (103.1°F), and the water
temperature is above 39.5°C (103.1°F)
when the system is in Manual Control
mode.
Patient temperature is not
automatically controlled while in
Manual Control mode
Decrease Manual Mode water target temperature to a setting that is ≤ 38.5°C
(101.3°F).
See
VI. Operation Guide–Therapy Settings–Manual Control
for further
instructions.
13 Patient Temperature
1 Low
Low Priority Alarm
The Patient Temperature 1 reading
is below 31°C (87.8°F), and the water
temperature is below 31°C (87.8°F)
when the system is in Manual Control
mode.
Patient temperature is not
automatically controlled while in
Manual Control mode
Increase Manual Mode water target temperature to a setting that is ≥ 32°C
(89.6°F).
See
VI. Operation Guide–Therapy Settings–Manual Control
for further
instructions.