The current Service Manual can be found at www.medivance.com/manuals
Fill Reservoir
Approximately four liters of sterile water will be required to fill
the reservoir at initial installation. Fill the reservoir with sterile
water only. When filling the Control Module during initial
installation or when completely empty, add one vial of Arctic
Sun™ Temperature Management System Cleaning Solution
to the sterile water. It is recommended to add the vial when
filling with the second liter of sterile water.
1. From the Patient
Therapy Selection
screen, select the
button next to either Normothermia or Hypothermia
under the New Patient heading. Select any pad type to
continue.
2. From the
Hypothermia
or
Normothermia
therapy
screen, press the
Fill Reservoir
button.
3. The
Fill Reservoir
screen will appear. Follow the
directions on the screen.
4. The filling process will automatically stop when the
reservoir is full. Continue to replace the bottles of sterile
water until the filling process stops.
5. When the
Fill Reservoir
process is complete, the screen
will close.
6. To stop the process early, press the
Stop
button.
Note
: If the filling cycle is stopped prior to completion,
the reservoir will not be full and may requiring
filling after fewer patient therapies have been
performed.
7. Press the
Cancel
button to close the screen.
Fig 1-6 Fill Reservoir Screen
Functional Verification
Certificates of Conformance for calibration, performance, and
electrical safety tests are included with the shipment of each
Arctic Sun™ Temperature Management System. To verify
the system will heat and cool properly, perform the following
functional verification procedure after initial setup and
installation of the Control Module.
1. Power On the Control Module.
2. From the
patient
Therapy Selection
screen, press
the button next to
Hypothermia
to display the
Hypothermia
therapy screen. Select the adult pad type
to continue.
Fig 1-7 Manual Control Settings Screen
3. From the
Hypothermia
therapy screen, press the
Manual Control
button to open the
Manual Contro
l
window. If the Manual Control button is not visible,
select the Adjust button at the bottom of the screen
and select the
More
button to access the therapy
settings screen. Enable Manual Control by pressing the
up arrow once and save your settings. Press close to
close the settings window.
4. From the therapy screen, press the
Manual Control
button. Use the Up and Down arrows to set the
Manual Control water target temperature to 40°C and
the duration to 30 minutes.
5. Press the
Start
button to initiate
Manual Control
.
Allow at least 3 minutes for the system to stabilize.
6. Monitor the flow rate and water temperature in the
System
status
area on the
Hypothermia
therapy
screen.
7. Verify that the flow rate reaches at least 1.5 liters/
minute.
8. Verify that the water temperature increases to at least
30°C.
9. Set the Manual Control water target temperature to
4°C and the duration to 30 minutes.
10. Monitor the flow rate and water temperature in the
System
status area of the
Hypothermia
therapy
screen. Verify that the water temperature drops to 6°C
or colder.
11. Press the
Stop
button to stop
Manual Control
.
12. Press the
Cancel
button to close the
Manual Control
window.
13. Before placing the device into clinical use, it is
recommended to disable Manual Control unless
requested by clinical users. Select the
Adjust
button at
the bottom of the therapy screen and select the
More
button to access the therapy settings screen. Disable
Manual Control and save your settings to close the
window.
14. Power Off the Control Module.