10
Service Manual
Inspect Screen Protector
The Control Panel’s touchscreen is supplied with a
disposable screen protector. If it becomes damaged, it can
be removed by lifting the edge and carefully peeling it from
the screen. To ensure dust and particulates are removed,
clean the touchscreen using isopropyl alcohol. Remove the
blue liner from the screen protector. Then carefully apply the
protector to the screen with the liner side down.
Inspect Fluid Delivery Line
1. Power On the system
2. From the patient
Therapy Selection
screen press
the Hypothermia button to display the Hypothermia
therapy screen.
3. From the Hypothermia therapy screen, press the
Manual Control button to open the Manual Control
window.
4. Set the Manual Control water target temperature to
28°C and the duration to 30 minutes.
5. Connect a shunt to a set of fluid delivery line ports.
6. Press the
System Access
button on the Therapy screen
then press the
Diagnostics
button. Verify that inlet
pressure is -7 psi ± 0.2.
7. Repeat on all valves. If inlet pressure is out of range,
replace the two valves that the shunt is connected to.
8. Ensure that the shunt is removed before device is put
back in service.
Inspect Manifold O-Rings
Inspect the O-rings in the manifold. Ensure no cuts or tears
are present and that the rings make an intact seal with the
Fluid Delivery Line.
System Diagnostics
Select the
System Access
button on the
Normothermia
therapy or
Hypothermia
therapy screen to access
System
Diagnostics
in order to verify pressures and flow rates
through the system.
Calibration
To perform a calibration on the Arctic Sun
™
Temperature
Management System, press the
Advanced Setup
button
on the
Therapy Selection Screen
. Press the
Start
button
and follow the on-screen directions. Refer to Chapter 9 for
additional instructions.
Functions
• Download Patient Data
• Connectivity Settings
• Calibration
• Total Drain
• Save All Settings As Default
Location / Time Settings
• Number Format
• Current Time
• Date Format
• Current Date
Information
• Software Versions
• Last Calibration Date
• Next Calibration Due
Fig 5-1 Advanced Setup Screen
Chapter 5 - Advanced Setup
Use the
Advanced Setup
screen to view the current settings
and modify the settings for the following parameters. To
modify any parameter setting, press the button to the right
of the parameter.
To access the Advanced Setup screen:
1. Press
Advanced Setup
button on the patient
Therapy
Selection
screen.
2. The Advanced Setup screen will be displayed.
Note:
The
Advanced Setup
button on the Patient Therapy
screen will be available when the unit is Powered On.
The
Advanced Setup
button will not be available
once patient therapy has been started. The
Advanced
Setup
button can only be redisplayed after cycling the
power.