24
Service Manual
Alarm
Problem
Solution
53 Prolonged Cold Water
Exposure
Medium Priority Alarm
The circulating water temperature
has been below 10°C (50°F) for
a prolonged period of time. The
extended period of cold water alarm
has been issued 11 times. The alarm
was first issued after the system
sensed that the water temperature
was below 10°C (50°F) for 8 of 10
hours. The alarm was then issued
an additional 10 times every
1 hour because the situation was not
resolved.
Prolonged cold water exposure
may increase the risk for skin injury.
Assess patient’s skin underneath the
ArcticGel™ pads.
1. Verify the custom parameters.
Patient target is set to the correct temperature.
See
VI. Operation Guide–Therapy Screen–Hypothermia
Therapy–Cooling
or
Normothermia Therapy – Control Patient
for
instructions on setting patient target temperature.
One of the automatic patient control modes (e.g. Control Patient or
Cooling) is activated. (The system is not in Manual Control mode.)
The high water temperature limit and low water temperature limits
are set correctly. See
VI. Operation Guide–Therapy Settings–High
Water Limit
or
Low Water Limit
.
2. Verify Pad Sizing and Coverage
A full set of four ArcticGel™ pads of the appropriate size for the
patient applied to the patient.
For patients > 100 kg (220 lbs), 1 or 2 Universal pads are added as
required for adequate coverage.
The pads are well-adhered to the patient.
3. Verify System Performance
Water flow rate should be greater than the minimum flow rate
requirements specified in the ArcticGel™ pads IFU at least 1 hour
continuous use.
See
VIII. Troubleshooting-Water–Low Water Flow
for
troubleshooting instructions.
The patient temperature probe is properly placed and providing an
accurate and stable temperature.
4. Verify Shivering Control
If the device settings, pad sizing and system performance is
correct and patient target temperature is still not reached and/or
water temperature remains below 10˚C (50˚F), then the patient is
generating excessive heat, most likely from shivering which may or
may not be visible.
Consider administration of additional medication for shivering
control, adequate for the patient weight and magnitude of
shivering: follow physician orders, institutional TTM protocol and
current medical guidelines.
Evaluate patient response to medication.
5. Make Clinical Decision
If all of the above considerations have been addressed and the
patient still has not reached target temperature the physician
and nursing staff make a clinical decision to limit the cold water
exposure:
Increase low water temperature limit. See
VI. Operation Guide–
Therapy Settings–Low Water Limit
.
Set the patient target temperature to the lowest patient
temperature achieved. See
VI. Operation Guide–Therapy
Screen–Hypothermia Therapy–Cooling
or
Normothermia
Therapy – Control Patient
for instructions on setting patient target
temperature.
Discontinue cooling therapy.
60 Non-Recoverable
System Error
Medium Priority Alarm
Control processor and Monitor
processor start up synchronization fault.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
61 Non-Recoverable
System Error
Medium Priority Alarm
Control processor parameter memory
fault.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
62 Non-Recoverable
System Error
Medium Priority Alarm
Monitor processor parameter memory
fault.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
64 Non-Recoverable
System Error
Medium Priority Alarm
Unable to enable pump power
(Control processor).
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
65 Non-Recoverable
System Error
Medium Priority Alarm
Unable to enable pump power
(Monitor processor).
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
66 Non-Recoverable
System Error
Medium Priority Alarm
Unable to disable pump power
(Control processor).
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.