40
Service Manual
8.10 Replace the USB to RS232 Converter (T.I.M.)
1. Disconnect the RS232 cable from the back of the
USB to RS232 converter (T.I.M.).
2. Detach the USB to RS232 converter (T.I.M.) from
the top of the card cage.
4. Disconnect the RS232 cable from the back of the
USB to RS232 converter (T.I.M.) and the USB cable
from connector panel bracket then remove the
connector panel bracket assembly from the device.
5. Remove the panel ESD board with blue temp
connector PCB cable from the connector panel per
step 8.8 and install on the new Connection Bracket
Assembly.
6. Reconnect and route all wiring, reference (figure
8-25) for placement of AC Breaker harness on the
back of the power switch.
Note
: Orientation of AC Breaker Harness Connections:
•
Brown wrapped with orange stays with orange
•
Brown wrapped with purple stays with purple
7. Assemble the connector panel bracket assembly to
the device and connect the blue temp connector
PCB cable to the isolation board.
Fig 8-25 Placement of AC Breaker Harness Connections
Fig 8-26 Connector Panel Bracket Assembly
3. Assemble the new USB to RS232 converter
(T.I.M.) to the top of the card cage with the USB
connection towards the connector panel (see figure
8-27).
4. Reconnect the RS232 cable to the back of the USB
to RS232 converter (T.I.M.).
Fig 8-27 Orientation of T.I.M. on top of the card cage
Fig 8-28 USB to RS232 Converter (T.I.M.)
8.11 Replace the Level Sensor
Tools and Supplies required:
• 7/16” (11mm) nut driver
• Wire cutters
1. Disconnect the gray Level Sensor harness from the
I/O Circuit Card. Remove the cable tie securing the
harness to the bracket.
2. Using a 7/16” (11mm) nut driver, remove the screw
securing the level sensor bracket.
3. Remove the level sensor along with the plastic
sheath from the tank.
4. Insert the new level sensor into the tank, ensure the
curve of the tube is facing the bottom of the tank.
5. Reinstall the bracket, reroute and connect the gray
Level Sensor harness to the I/O Circuit Card.