Surface Disinfection
Disinfect the Arctic Sun™ Temperature Management System
immediately following cleaning. Follow all manufacturers’
guidelines for the pre-saturated germicidal wipes, including
wearing gloves and other personal protective equipment
when dispensing and using the wipes.
Manual Disinfection:
• Disinfection Materials:
Required
: Pre-saturated germicidal wipes (active
ingredient – alcohol)
Validated
: Super Sani-Cloth Germicidal Disposable Wipe
• Disinfection
1. Using pre-saturated germicidal wipes, disinfect all
external surfaces of the Arctic Sun™ Temperature
Management System.
a) Ensure that all surfaces are dampened, including all
seams of the device.
b) Use as many clean towelettes as necessary to
ensure device is completely dampened with
disinfectant.
c) Thoroughly wipe the following with additional
clean towelettes – fluid delivery line, power cable,
temperature cables, USB cable, RS232 cable.
2. Allow the treated surfaces to remain wet for the
manufacturer’s specified contact time.
3. Wipe each surface of the device with a lint-free cloth
premoistened with sterile water.
4. Let air dry.
5. Reattach cables and hoses.
Visual Inspection and Return to Use
1. After cleaning and disinfecting the device per
the instruction provided, examine the device for
cleanliness. If visible soil remains, repeat manual
cleaning instructions.
2. Inspect cables and hoses for signs of damage.
3. Inspect external shell for discoloration or cracking.
4. Inspect labels for legibility.
5. Do not use the device if it has failed visual inspection
for soil after multiple cleaning attempts, or if there
is visible damage. Contact Customer Support for
additional recommendations.
6. Store the device in a dry and clean environment when
not in use.
Note:
The device has been validated for exterior surface
cleaning and disinfection per the instructions provided.
Any deviation from the recommended parameters must
be validated by the user. Interior surface of the device
(including deep crevices) are not addressed as part of this
cleaning and disinfection protocol.
To prevent possible discoloration, do not use iodine-based
solutions, such as Betadine, on any part of the machine.
For questions on device cleaning and disinfection of
external surfaces contact BD Customer Support.
Inspect Connectors and Cables
Inspect the patient temperature cable(s) and power cord
for integrity. Ensure temperature cables are properly strain
relieved. Ensure power cord bracket is secure.
Condenser
• A dirty chiller condenser will significantly reduce the
cooling capacity of the Control Module.
• To clean the condenser, wipe the dust from the exterior
grill using a soft cloth. Depending on the quality of your
institution’s air, periodically remove the back cover and
vacuum or brush the condenser fins. At a minimum the
condenser fins should be cleaned annually. Maintenance
activities should be performed by qualified personnel.
Device Inspection
• Periodically inspect the external areas of the device for
damaged, loose or missing parts, and frayed or twisted
power cords and cables.
• Discontinue using the device displaying one or more of
the above conditions until the problem is corrected and
has been verified to be operating correctly.
Replenish Internal Solution
Contact Customer Service to order internal Arctic Sun
Cleaning Solution.
To replenish the internal Arctic Sun Cleaning Solution:
1. Drain the reservoir.
• Turn Control Module Power Off.
• Attach the drain line to the two drain ports on the
back of the Control Module. Place the end of the drain
line into a container. The water will passively drain
into the container.
2. Refill the reservoir.
• Connect the fill tube.
• From the Hypothermia therapy screen or the
Normothermia therapy screen, press the Fill Reservoir
button.
• The Fill Reservoir screen will appear. Follow the
directions on the screen.
• Add one vial of Arctic Sun™ Temperature
Management System cleaning solution to the
second liter of sterile water.
• The filling process will automatically stop when the
reservoir is full. Continue to replace the bottles of
sterile water until the filling process stops.
• When the Fill Reservoir process is complete, the screen
will close.
Note:
The reservoir level sensor requires a conductive fluid to
operate properly. Filling the reservoir without using the
Arctic Sun Cleaning Solution may result in overfilling
the reservoir.