MEDIANA
HEART
ON
AED A16
Automated External Defibrillator
3
3
Cautions
Caution statements identify conditions or practices that could result in damage to
the equipment or other property.
The AED may not operate properly if it is operated or stored at conditions
outside the ranges stated in this manual.
The AED was designed to be sturdy and reliable for many different use
conditions. However, handling the AED too roughly can damage it or its
accessories and will invalidate the warranty. Check the AED and
accessories regularly for damage, according to directions.
Before delivering a shock, it is important to disconnect the patient from
non-defibrillation protected electronic devices, such as blood-flow
meters, that may not incorporate defibrillation protection. In addition,
make sure the pads are not in contact with metal objects such as a bed
frame or stretcher.
The pads pouch shall not be opened until immediately prior to use.
Do not use or place the AED in service until you have read the AED
Operator’s manual.
Do not use or stack the AED with other equipment. If the AED is used or
stacked with other equipment, verify proper operation prior to use.
Handling or transporting the patient during ECG analysis can cause
incorrect or delayed diagnosis. If the AED gives a SHOCK ADVISED
prompt during such handling or transport, stop the vehicle and keep the
patient as still as possible for at least 15 seconds before pressing the
Shock button, to allow the AED to reconfirm the rhythm analysis.
Periodic checks of this AED must be undertaken to ensure among other
things that the AED is not damaged in any way.
The pads are a single use item and must be replaced after each use or if
pouch that seals pads has been broken/compromised in any way. If
damage is suspected the pads must be replaced immediately.
Do not use training pads with this AED.
Carefully observe pacemaker patients. Patient history and physical
examination are important in determining the presence of implanted
pacemaker. Patient pacemakers may reduce the sensitivity of the AED
analysis and errors in detecting shockable rhythms.
If the pads are attached to the chest firmly, the AED can analyze the exact
ECG and prevent the skin burns. But if the pads are overlapped on the
patient chest, the pads will not deliver defibrillation energy properly.
If the voice recording option is applied, the HeartOn A16
records the
surrounding sound. The voice prompt,
“
Audio recording in progress
”
, is
played.