![Mediana A16M-DF-0E Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/mediana/a16m-df-0e/a16m-df-0e_operators-manual_1760607049.webp)
MEDIANA
HEART
ON
AED A16
Automated External Defibrillator
45
45
TROUBLESHOOTING
If you are uncertain about the accuracy of any measurement, check the
patient’s vital signs by alternate means; then make sure the
AED is
functioning correctly.
To reduce the risk of electrical shock, do not attempt to remove the cover
under any circumstances. There are no operator serviceable components
and only a qualified technician should service the AED.
Please contact qualified service personnel or your local supplier if
unintentional equipment changes or attacks from outside is found.
General
If the AED detects an error, it can display the
“
X
”
on the status indicator. Contact qualified
service personnel or your local supplier. Before calling to qualified service personnel or
your local supplier, make sure it meets environmental conditions provided in the manual
as temperature, humidity, altitude and so on.
Corrective Action
Check used by date the pads. Change the pads if use by date has been exceeded.
Check shelf life or standby life of the battery. Change the battery if the shelf life or
standby life has been exceeded.
Following is a list of possible errors and suggestions for corrective action.
If the status indicator is still not displaying
“
X
”
or a warning message is heard when the
AED is turned on or if for any reason, you have suspicions that your AED is not working
correctly contact qualified service personnel or your local supplier or Mediana directly for
support. (
)
1. There is no response to the pushing the power button of the AED.
A main board may be malfunctioned. Notify qualified service personnel or your
local supplier to check and replace the main board.
The battery may be missing or discharged. If the battery is missing, insert the
battery (See Battery Operation section). If the battery is discharged, change the
battery. (See Maintenance section)
2. The beep tones do not sound during the operation.
Do not use the AED; contact qualified service personnel or your local supplier.
3. The beep tones sound but voice does not function properly.
Do not use the AED; contact qualified service personnel or your local supplier.
4. The voice p
rompt “Plug in pads. Insert connector firmly”
.
Reconnect the pad connector with pad socket firmly or replace the pad.
5. The action icon does not flash.
Do not use the AED, contact qualified service personnel or your local supplier.
6. The voice prompt is unclearly heard.
Do not use the AED, contact qualified service personnel or your local supplier.
7. The battery status does not indicate 3 bar despite of replacing new battery.
If the battery status still not displayed 3 bar despite of replacing new battery, do not
use the AED and contact qualified service personnel or your local supplier.
8. The pads status icon is displayed despite of replacing new pads.
If the pads status icon is displayed despite of replacing new pads, do not use the
AED and contact qualified service personnel or your local supplier.
Содержание A16M-DF-0E
Страница 8: ...MEDIANA HEARTON AED A16 Automated External Defibrillator 4...
Страница 16: ...MEDIANA HEARTON AED A16 Automated External Defibrillator 12...
Страница 28: ...MEDIANA HEARTON AED A16 Automated External Defibrillator 24...
Страница 48: ...MEDIANA HEARTON AED A16 Automated External Defibrillator 44...