Pre Defibrillation Action
Prior to using the AED, it is advised to perform the following checks and actions in order
to prepare the patient.
Remove clothes to expose bare chest.
If excessively hairy shave hair from areas to which defibrillation pads are to be
applied.
Ensure that the patient chest is dry. If necessary, dry chest area.
Operating the AED
The Pediatric mode must be used on patients between 1 and 8 years or
less than 25 kg (55lb).
If the pads placement is inappropriate, the AED could harm the patient.
To place the accurate position, must follow the voice prompt and action
icon. When pads placement is inappropriate, treatment could not work,
analysis could be incorrect, shock or no shock decision advisory could
be inappropriate, or shock could burn the patient
’
s skin.
Do not place pads near the generator of an internal pacemaker. The
analyzing heart rhythm of patient who is implanted pacemaker could
inaccurate or the pacemaker might be damaged by defibrillator
discharges.
Do not perform chest compressions (CPR) through electrodes. These
actions may damage the electrode pads cause the AED to function
improperly.
Always apply pads to flat areas of skin. Avoid application over folds of
skin such as those underneath the breast or on obese patients.
Excessive hair, poor adhesion, or air under pads may produce burns or
ineffective energy transfer.
To apply the pads to patient chest properly, shave hair from areas which
defibrillation pads are to be applied if necessary.
Always check the use by date on the pads and do not use the pads if the
packaging has been previously opened. If the excessively dry pads are
attached, AED may interpret as a condition that the pads are not attached
to the patient.
Touching the patient during the analysis phase of treatment can cause
interference with the diagnostic process. Avoid contact with the patient
while analysis is being carried out. The AED will instruct you by voice
prompt when it is safe to touch the patient.
The AED delivers shocks which can cause serious harm to operators and
bystanders. Caution must be taken to ensure no-one is in contact with
the patient when a shock is delivered.
Make sure no one is touching the patient before you press the Shock
button (in case of the Semi-Auto). Loudly announce,
“
Stand back! Do not
touch the patient.
”
And look down the entire length of the patent to
ensure there is no contact with a bystander or conductive surface before
pressing the Shock button.
Make sure no one is touching the patient before the shock is
automatically activated (in case of the Fully-Auto).
“
Shock will be
delivered. Do not touch the patient. 3, 2, 1.
”
When this voice is played, the
electric shock is transmitted automatically, so look down the entire
length of the patient to ensure there is no contact with a bystander or
conductive surface.
Do not touch the electrode surfaces, the patient, or any conductive