Table 9.
Immunity to proximity fields from RF wireless communications equipment (IEC
60601-1-2)
Test
frequency
(MHz)
Band
a
(MHz)
Service
a
Modulation
b
Maximum
power
(W)
Distanc
e
(m)
Immunity
test level
(V/m)
385
360
–
390
TETRA
400
Pulse
modulation
b
18 Hz
1.8
0.3
27
450
430
–
470
GMRS
460, FRS
460
FM
c
± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
745
780
704
–
787
LTE Band
13, 17
Pulse
modulation
b
217 Hz
0.2
0.3
9
810
870
930
800
–
960
GSM
800/900,
TETRA
800,
iDEN 820,
CDMA 850,
LTE Band
5
Pulse
modulation
b
18 Hz
2
0.3
28
1720
1845
1970
1700
–
1990
GSM 1800;
CDMA
1900;
GSM 1900;
DECT;
LTE Band
1, 3, 4, 25;
UMTS
Pulse
modulation
b
217 Hz
2
0.3
28
2450
2400
–
2570
Bluetooth,
WLAN,
802.11
b/g/n,
RFID 2450,
LTE Band
7
Pulse
modulation
b
217 Hz
2
0.3
28
5240
5500
5785
5100
–
5800
WLAN
802.11 a/n
Pulse
modulation
b
217 Hz
0.2
0.3
9
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the
transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m
test distance is permitted by IEC 61000-4-3.
a
For some services, only the uplink frequencies are included.
b
The carrier shall be modulated using a 50 % duty cycle square wave signal.
c
As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because
while it does not represent actual modulation, it would be worst case.
Table 10. Cables (IEC60601-1-2)
Cables and Sensors
Maximum
Length
Complies with
Pads cable
1.8 m
-RF emissions, CISPR 11, Class B/ Group 1
-Electrostatic discharge (ESD), IEC 61000-4-2
-Radiated RF, IEC 61000-4-3
-Power frequency Magnetic field, IEC 61000-4-8