If the pad is not inserted, you will hear the voice prompt:
“Plug in pads. Insert connector firmly
.
”
Figure 9. Pads disconnect icon (Chest type
∙
A16-DS, A16-DF)
Figure 10. Pads disconnect icon (Icon type
∙
A16-GS, A16-GF)
If the AED that is activated Pads Quality function and there is a problem with the pads,
you will hear the voice prompt:
“
Replace pads.
”
Note: If the pad connector is not connected in any step except for Step 4 (CPR), the AED
will move to Pad connector disconnected icon and the voice prompt
“
Plug in pads.
Insert connector firmly.
”
is emitted.
Note: If the pad is disconnected during Step 2 (ECG analysis) or Step 3 (delivering
electric shock), notify the user that the electric shock has been canceled, and then
the voice prompt
“
Plug in pads. Insert connector firmly.
”
is emitted. However, if the
electric shock is canceled more than 3 times before the completion of Step 3
(delivering electric shock), perform Step 4 (CPR).
7. Verify the AED up to
‘
6.
’
which is activated normally and follow voice prompt and action
icon. The red LED will flash near the relevant action icon.